Farklı elektronik apeks bulucuların kök perforasyonlarının tespitinde etkili olabildikleri minimum perforasyon çapının belirlenmesi
Accuracy of different electronic apex locators in determination of minimum root perforation diameter
- Tez No: 687830
- Danışmanlar: PROF. DR. ALPER KUŞTARCI
- Tez Türü: Diş Hekimliği Uzmanlık
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Elektronik apeks bulucu, perforasyon, kök kanalı, Dental aygıtlar, Dental pulpa boşluğu, Diş apeksi, Diş kökü, Electronic apex locator, perforation, root canal, Dental instruments, Dental pulp cavity, Tooth apex, Tooth root
- Yıl: 2021
- Dil: Türkçe
- Üniversite: Akdeniz Üniversitesi
- Enstitü: Diş Hekimliği Fakültesi
- Ana Bilim Dalı: Endodonti Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Amaç: Bu çalışmanın amacı, 6 farklı elektronik apeks bulucu (EAB) olan Raypex 6 (VDW, Münih, Almanya), Propex Pixi (PIXI, Dentsply Maillefer, Ballaigues, İsviçre), Dentaport ZX (J. Morita, Irvine, Kaliforniya, ABD), Apex ID (SybronEndo, Orange, Kaliforniya, ABD), Propex II (PRO; Dentsply Maillefer, Ballaigues, İsviçre) ve Dr.'s Finder NEO (Good Doctors, Incheon, Kore) ile EAB'lerin perforasyon tespitinde etkili olabildikleri minimum perforasyon çapını belirlemektir. Yöntem: 100 adet çekilmiş mandibular premolar diş her grupta 20 adet diş olacak şekilde 5 farklı çaptaki elmas frezle (1,25 mm, 1,0 mm, 0,75 mm, 0,5 mm, 0,25 mm) apikalden 5 mm yukarıda olacak şekilde perfore edildi. Perforasyon alanına kadar olan gerçek uzunluk (GU) ölçüldükten sonra #40 no K tipi eğe kullanılarak dişler aljinat model içine gömülerek elektronik uzunluk (EU) ölçüldü. Her diş için EU, GU'dan çıkarılarak aradaki fark hesaplandı. İstatiksel analiz Two-way Anova testi kullanılarak yapıldı (P<0,05). Bulgular: Çalışmada kullanılan EAB'ler arasında 1,25 ve 1,00 mm perforasyon çapları için anlamlı fark bulunmadı (p>0,05). Perforasyon çapının 0,75 mm olduğu dişlerde Propex Pixi ile Dentaport ZX arasında anlamlı fark bulundu (p<0,05). Sonuç: Test edilen tüm EAB'ler, 0,5 mm ve 0,25 mm çapları için perforasyon alanını tespit edememiş olsa da 1,25 mm, 1 mm, 0,75 mm çapları için simüle edilmiş kök perforasyonlarının belirlenmesinde oldukça başarılı olmuşlardır.
Özet (Çeviri)
Objective: The aim of this study was to demonstrate the detection of minimum root canal perforation diameter with Raypex 6 (VDW, Munich, Germany), Propex Pixi (Dentsply Maillefer, Ballaigues, Switzerland) Dentaport ZX (J. Morita, Irvine, California, USA), Apex ID (SybronEndo, Orange, CA, USA), Propex II (PRO; Dentsply Maillefer, Ballaigues, Switzerland), Dr's Finder NEO (Good Doctors, Incheon, Korea) electronic apex locators (EALs). Materials and Methods: One-hundred extracted, single-rooted human teeth were artificially perforated using burs with 5 different diameters (1.25 mm, 1.0 mm, 0.75 mm, 0.5 mm, 0.25 mm) in 5 mm above the apex. Each group consisted of 20 teeth. The actual canal lengths (AL) up to the perforation site were measured, and then the teeth were embedded in an alginate mold. Perforations were electronically measured using a size 40 K-file by each EAL. For each tooth, AL was subtracted from the electronic canal length (EL) of the perforation. Statistical analyses were performed using Two-way Anova test at a significance level of 0.05. Results: There was no significant difference among all EALs used in this study at 1.25 and 1.00 mm perforation diameters (p> .05). A significant difference was found between Propex pixi and Dentaport ZX in teeth with a perforation diameter of 0.75 mm (p <.05). Conclusion: Although all EALs tested could not detect the perforations for diameters of 0.5 mm and 0.25 mm, they were highly successful in determination of simulated root perforations for diameters of 1.25 mm, 1 mm, 0.75 mm. *
Benzer Tezler
- Kök kanal perforasyonlarında farklı kanal içi ortamların apeks bulucuların doğruluğuna etkisi
The effect of different intracanal conditions on the accuracyof apex locators in the presence of root canal perforations
ASENA OKUR
Diş Hekimliği Uzmanlık
Türkçe
2021
Diş HekimliğiErciyes ÜniversitesiEndodonti Ana Bilim Dalı
DOÇ. DR. TUĞRUL ASLAN
- Farklı apeks bulucu sistemlerin farklı irrigasyon solüsyonları varlığında kök perforasyonlarını belirlemedeki doğruluğunun değerlendirilmesi
Evaluation of different apex finder systems for determination of root perforations in presence of different i̇rrigation solutions
OĞUZ BURHAN ÇETİNKAYA
Diş Hekimliği Uzmanlık
Türkçe
2019
Diş HekimliğiGaziantep ÜniversitesiEndodonti Ana Bilim Dalı
DOÇ. UĞUR AYDIN
- Yapay olarak oluşturulmuş apikal kök rezorpsiyonlu dişlerde farklı elektronik apeks bulucuların çalışma uzunluğu saptama etkinliklerinin incelenmesi
Investigation of working length determination efficiency of different electronic apex locators in TEETH with simulated apical root resorption
DUYGU KÜLAHCI
- Üç farklı elektronik apeks bulucu, radyovizyografi ve konik ışınlı bilgisayarlı tomografinin süt dişi kök kanal tedavisinde çalışma boyu belirlenmesinde doğruluğunun değerlendirilmesi
The accuracy of three different electronic apex locators, radiovisiography and cone-beam computed tomography in determination of the working lenght measurement on primary teeth root canal treatment
MEVLÜT KAYABAŞI
Diş Hekimliği Uzmanlık
Türkçe
2019
Diş HekimliğiSivas Cumhuriyet ÜniversitesiÇocuk Diş Hekimliği Ana Bilim Dalı
DOÇ. DR. FATİH ÖZNURHAN
- Kök kanal tedavisinde kanal boylarının tespitinde kullanılan farklı apeks bulucuların doğruluğunun periapikal radyografi ve konik ışınlı bilgisayarlı tomografi ile değerlendirilmesi
Evaluation of the accuracy of different apex locator used in detection of canal length in root canal treatment with periapical radiography and cone beam computed tomography
KAZIM DELİCE
Diş Hekimliği Uzmanlık
Türkçe
2022
Diş HekimliğiOndokuz Mayıs ÜniversitesiEndodonti Ana Bilim Dalı
PROF. DR. HİKMET AYDEMİR
PROF. DR. ALİ KELEŞ