Ultrasonik pakimetri, speküler mikroskopi ve optik koherans tomografi ile ölçülen merkezi kornea kalınlık verilerinin karşılaştırılması
Comparison of central cornea thickness data measured by ultrasonic pacimetry, specular microscopy and optic coherance tomography
- Tez No: 639860
- Danışmanlar: YRD. DOÇ. DR. ÖZKAN SEVER
- Tez Türü: Tıpta Uzmanlık
- Konular: Göz Hastalıkları, Eye Diseases
- Anahtar Kelimeler: Santral Kornea Kalınlığı, Ultrasonik Pakimetre, CSO Speküler mikroskopi, Optik Koherans Tomografi, tekrarlanabilirlik, Tomografi-optik koherens, Central Korneal Thickness, Ultrasonic Pachymetry, CSO Specular Mikroscopy, Optical Koherens Tomography, repeatability, Tomography-optical coherence
- Yıl: 2017
- Dil: Türkçe
- Üniversite: Tekirdağ Namık Kemal Üniversitesi
- Enstitü: Tıp Fakültesi
- Ana Bilim Dalı: Göz Hastalıkları Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
AMAÇ: Ultrasonik pakimetri (Pacscan 300P, Sonomed, New York, ABD), CSO speküler mikroskop (SP-01, CSO, Floransa, İtalya) ve Cirrus optik koherans tomografiden (Model 4000, Carl Zeiss, Meditec, Dublin, CA, ABD) elde edilen merkezi kornea kalınlık verilerinin tekrarlanabilirliği ve karşılaştırılabilirliğini değerlendirmek. MATERYAL VE METHOD: Yaşları 20 ile 53 yıl arasında değişen 48 sağlıklı gönüllünün 96 gözünün merkezi kornea kalınlığı her bir cihaz ile üçer kez ölçüldü. Her bir cihazdan elde edilen 3 ölçümün tekrarlanabilirliği varyasyon katsayısı (%) ve sınıf içi korelasyon katsayısı ile değerlendirildi. Cihazlar arasındaki merkezi kornea kalınlık farkını karşılaştırmak için wilcoxon analizi kullanıldı. BULGULAR: Ultrasonik pakimetri, CSO optik pakimetri ve Cirrus optik koherans tomografi için varyasyon katsayıları, sırasıyla 6.5, 7.3 ve 6.2 iken sınıf içi korelasyon katsayıları sırasıyla 0.993, 0.968, 0.998 idi. Ortalama kornea kalınlığı sırasıyla 561.5, 555.7 ve 553.0 mikrondu. CSO optik pakimetri ve Cirrus optik koherans tomografiden elde edilen merkezi kornea kalınlık verileri ultrasonik pakimetriden elde edilen verilerden önemli derecede düşüktü (sırasıyla, p<0.001, p<0.001). SONUÇ: Optik pakimetri cihazları merkezi kornea kalınlığını standart ultrasonik pakimetreden daha ince ölçmektedir. CSO optik pakimetre cihazından elde edilen kornea kalınlık ölçümlerinin tekrarlanabilirliği diğer cihazlara göre daha düşüktür.
Özet (Çeviri)
PURPOSE: To evaluate the reproducibility and comparability of central corneal thickness data obtained from Ultrasonic pachymetry (Pacscan 300P, Sonomed, New York, ABD), CSO specular microscopy (SP-01, CSO, Florence, Italy) and Cirrus optical coherence tomography (Model 4000, Carl Zeiss, Meditec, Dublin, CA, ABD). MATERİALS AND METHODS: The central corneal thickness of 96 eyes of 48 healthy volunteers aged 20 to 53 years was measured three times with each device. The reproducibility of the 3 measurements obtained from each device was evaluated by the coefficient of variation (%) and the intraclass correlation coefficient. Wilcoxon analysis was used to compare the central corneal thickness difference between the devices. RESULTS: The coefficients of variation for ultrasound pachymetry, CSO optical pachymetry, and Cirrus optical coherence tomography were 6.5, 7.3, and 6.2, respectively, while the intraclass correlation coefficients were 0.993, 0.968, and 0.998, respectively. The mean corneal thickness was 561.5, 555.7 and 553.0 microns, respectively. Central corneal thickness obtained from CSO optical pachymetry and Cirrus optical coherence tomography was significantly lower than the data obtained from ultrasonic pachymetry (p <0.001, p <0.001, respectively). CONCLUSIONS: Optical pachymetry devices measure the thickness of the central cornea thinner than standard ultrasonic pachymetry. The reproducibility of the corneal thickness measurements obtained from the CSO optical pachymeter is lower than other devices.
Benzer Tezler
- Penetran keratoplasti sonrası kornea kalınlığının pakimetri ve ön segment optik kohorens tomografi (OKT) ile takibi
Follow-up and evaluation of corneal thickness after penetrating keratoplasty using pacimetry and anterior segment optical kohorens tomography (OCT)
YASEMİN DERE
Tıpta Uzmanlık
Türkçe
2014
Göz Hastalıklarıİstanbul ÜniversitesiGöz Hastalıkları Ana Bilim Dalı
PROF. DR. KORAY AKARÇAY
- Kornea ödemi tedavisinde sürfaktanın etkinliğinin değerlendirilmesi
Evaluation of the effectiveness of surfactant in the treatment of corneal edema
TUNCAY ARTUÇ
Tıpta Uzmanlık
Türkçe
2019
Göz HastalıklarıVan Yüzüncü Yıl ÜniversitesiGöz Hastalıkları Ana Bilim Dalı
DOÇ. DR. MUHAMMED BATUR
- Penetran keratoplasti yapılan hastaların santral korneal kalınlıklarının farklı cihazlarla karşılaştırılması
Comparison of central corneal thickness in patients undergoing penetrating keratoplasty using different devices
FATMA DİLARA ÖNAL ARTAN
Tıpta Uzmanlık
Türkçe
2026
Göz HastalıklarıManisa Celal Bayar ÜniversitesiGöz Hastalıkları Ana Bilim Dalı
DOÇ. DR. HÜSEYİN MAYALI
- Tavşan fakoemülsifikasyon modelinde ultrasonik titreşimlerinneden olduğu kornea endotel hasarına otolog serumun etkisi
The effect of autolog serum on corneal endothelial injury caused byultrasonic oscillation in rabbit facoemulsification model.
GÜLŞAH USTA
Tıpta Uzmanlık
Türkçe
2013
Göz HastalıklarıKırıkkale ÜniversitesiGöz Hastalıkları Ana Bilim Dalı
YRD. DOÇ. DR. REYHAN OĞUREL
- Tavşanlarda sefuroksı̇mı̇n farklı sulandırma solüsyonları ı̇le sulandırıldıktan sonra ı̇ntrakameral uygulama sonrası kornea üzerı̇ne lokal etkı̇lerı̇nı̇n değerlendı̇rı̇lmesı̇
Comparison of the local effects of different intracameral cefuroxime solutions on rabbit cornea
FERHAN YILMAZ
Tıpta Uzmanlık
Türkçe
2018
Göz HastalıklarıDokuz Eylül ÜniversitesiGöz Hastalıkları Ana Bilim Dalı
PROF. DR. AYŞE TÜLİN BERK