Nötropenik hastaların vücut sıcaklığı ölçümünde aksiller, temporal ve timpanik bölgelerden yapılan ölçümlerin karşılaştırılması
Comparison of measurement of body temperature fromaxillary temporal artery and tympanic membrane in neutropenic patients
- Tez No: 531936
- Danışmanlar: DOÇ. DR. DİLEK SARI
- Tez Türü: Yüksek Lisans
- Konular: Hemşirelik, Nursing
- Anahtar Kelimeler: nötropeni, vücut sıcaklığı, ateş takibi, elektronik aksiller termometre, temporal arter termometresi, timpanik membran termometresi, neutropenia, body temperature, electronic axillary thermometer, temporal artery thermometer, tympanic membrane thermometer
- Yıl: 2018
- Dil: Türkçe
- Üniversite: Ege Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Hemşirelik Esasları ve Yönetim Ana Bilim Dalı
- Bilim Dalı: Hemşirelik Esasları Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
Araştırma, nötropenik hastaların vücut sıcaklığı ölçümünde aksiller, temporal ve timpanik bölgelerden yapılan ölçümleri karşılaştırmak amacıyla yarı deneysel olarak yapılmıştır. Araştırma, Ağustos 2017- Mayıs 2018 tarihleri arasında İzmir Dr. Suat Seren Göğüs Hastalıkları ve Cerrahisi Eğitim ve Araştırma Hastanesi Göğüs Servislerinde (2, 3, 4 ve 5.göğüs servislerinde) yatarak tedavi gören erişkin hastalarda yapılmıştır. Araştırmanın örneklemini 18 yaş ve üstü, gönüllü, nötropeni tanısı ile yatarak tedavi gören erişkin hastalar (n=60-300 ölçüm) oluşturmuştur. Araştırmada veriler, Timpanik Membran Termometresi, Temporal Arter Termometresi, Elektronik Aksiller Termometre ve Hasta Bilgi Formu kullanılarak toplanmıştır. Verilerin toplanmasında, araştırmaya dahil olma ölçütlerine uygun hastalar belirlendikten sonra araştırma hakkında bilgi verilmiştir, gönüllü olan hastalardan bilgilendirilmiş onam alınmıştır. Hastalarla ilgili tanıtıcı bilgiler Hasta Bilgi Formu'na kaydedildikten sonra, aksiller, temporal ve timpanik bölgelerden yapılan vücut sıcaklığı ölçümleri Hasta Bilgi Formu'na kaydedilmiştir. Ölçümlere aksiller bölgeye elektronik aksiller termometrenin yerleştirilmesiyle başlanmış ve elektronik aksiller termometrenin ölçüm sonucu için sinyal vermesini beklerken sırasıyla timpanik ve temporal bölgelerden ölçümler yapılmıştır. Ölçümler arasındaki süre 10 saniyeyi geçmeyecek şekilde planlanmıştır. Verilerin analizinde IBM SPSS Statistics 21.0 (IBM SPSS Statistics for Windows, Version 21.0. Armonk, NY: IBM Corp.) programı kullanılmıştır. Verilerin istatistiksel analizlerinde frekans, yüzde, ortalama, standart sapma hesaplanmıştır. Ayrıca Bland Altman yöntemi ile verilere ilişkin grafikler oluşturulmuştur. Sonuçlar %95 güven aralığında, anlamlılık p<0.0001 düzeyinde değerlendirilmiştir. Araştırmanın yapılabilmesi için Ege Üniversitesi Tıp Fakültesi Klinik Araştırmalar Etik Kurulu'ndan ve İzmir Dr. Suat Seren Göğüs Hastalıkları ve Cerrahisi Eğitim ve Araştırma Hastanesinden örneklem kapsamına alınan her bir hastadan yazılı izin alınmıştır. Araştırmanın sonucunda ise, Araştırma kapsamına alınan hastaların, yaşları 50 ile 77 arasında değişmekte olup (X̄=61.96±7.18) %45' inin 60-69 yaş grubunda olduğu, % 91.7'sinin erkek olduğu saptanmıştır. Yapılan ölçümlerin saçılım grafikleri incelendiğinde, elektronik aksiller termometre ve kızılötesi temporal arter termometresi ile yapılan ölçümlerin timpanik membran termometresi ile yapılan ölçümlerden daha düşük olduğu saptanmıştır. Yapılan tüm ölçümlerde her termometrenin kendi içinde uyumlu olduğu saptanmıştır. Elektronik aksiller termometre ve timpanik membran termometresi ile yapılan tüm vücut sıcaklığı ölçümleri Bland-Altman grafiği incelendiğinde, ölçümler arasındaki ortalama fark 1.2730 oC olduğu aralarında ileri derecede anlamlı farklılık olduğu (p<0.0001) tespit edilmiştir. Kızılötesi temporal arter termometresi ve timpanik membran termometresi ile yapılan tüm vücut sıcaklığı ölçümleri Bland-Altman grafiği incelendiğinde, ölçümler arasındaki ortalama fark 0.4153 oC olduğu, aralarında ileri derecede anlamlı farklılık olduğu (p<0.0001) tespit edilmiştir. Elektronik aksiller termometre ve timpanik membran termometresi ile yapılan tüm vücut sıcaklığı ölçümleri Bland-Altman grafiği (X̄=1.27±0.54) ve kızılötesi temporal arter termometresi ve timpanik membran termometresi ile yapılan tüm vücut sıcaklığı ölçümleri Bland-Altman grafiği (X̄=0.41±0.43) incelendiğinde; temporal arter termometresi ile yapılan vücut sıcaklığı ölçümlerinin timpanik membran termometresi ile yapılan vücut sıcaklığı ölçümlerine daha yakın olduğu bulunmuştur.
Özet (Çeviri)
Research was quasi -experimental in order to compare of measurement of body temperature from axillary temporal artery and tympanic membrane in neutropenic patients. Research was carried out with adult patients who receiving inpatient treatment in Izmir Dr. Suat Seren Chest Diseases and Surgery Training and Research Hospital Chest Services (2, 3, 4 and 5th chest services) between August 2017- May 2018. The sampling of research consisted of volunteer adult patients (18 years old or above) who were receiving therapy as in-patient with a diagnosis of neutropenia (n=60- 300 measurement). In research data were collected by using Tympanic Membrane Thermometer, Temporal Artery Thermometer, Electronic Axillary Thermometer and Patient Information Form. During data collection, after patients that met inclusion criteria were identified, they were given information related to research and those that were volunteered to participate have provided informed consent. Introductory information about the patients after recording on the Patient Information Form, body temperature measurements from axillary, temporal and tympanic were recorded on the Patient Information Form. Measurements were initiated by the placement of the axillary thermometer in the axillary and measurements were taken from the tympanic and temporal respectively while the electronic axillary waited for the thermometer to signal for the measurement result. The time between measurements is planned not to exceed 10 seconds. IBM SPSS Statistics 21.0 (IBM SPSS Statistics for Windows, Version 21.0. Armonk, NY: IBM Corp.) was used in the analysis of data. In statistical analyses of data, the frequency, percentage, average and standard deviation were calculated. In addition, graphics related to data were drawn through Bland Altman method. Results were assessed at 95 % confidence interval and at p<0.0001 significance level. In order for the research can be performed, Ethical Board of Ege University Faculty of Medicine Clinical Research Ethics Committee and İzmir Dr. Suat Seren Chest Diseases and Surgery Training and Research Hospital's written approvals were obtained and each patient included in the scope of research has provided written informed consent. At the end of research it was identified that ages of the patients included in the scope of research were varied between 50 and 77 years (X̄=61.96±7.18); 45% were in 60-69 years age group and 91.7% were males. When the scatter plots of the measurements were examined, it was determined that the measurements made with the electronic axillary thermometer and the infrared temporal artery thermometer were lower than with the tympanic membrane thermometer. It has been determined that each thermometer is compatible with itself in all measurements made. When the Bland-Altman graph of whole body temperature measurements with electronic axillary thermometer and tympanic membrane thermometer was examined, the mean difference between the measurements was 1.2730 ° C and there was a significant difference between them (p <0.0001). When the Bland-Altman graph of whole body temperature measurements with infrared temporal artery thermometer and tympanic membrane thermometer was examined, the mean difference between the measurements was 0.4153 ° C and there was a significant difference between them (p <0.0001). When the Bland-Altman graph of whole body temperature measurements with electronic axillary thermometer and tympanic membrane thermometer (X̄=1.27±0.54) and the Bland-Altman graph of whole body temperature measurements with infrared temporal artery thermometer and tympanic membrane thermometer (X̄=0.41±0.43) were examined, the body temperature measurements with the temporal artery thermometer were found to be closer to the body temperature measurements with the tympanic membrane thermometer.
Benzer Tezler
- 1-36 ay yaş grubu çocuklarda temporal arter termometresi ile vücut ısısı ölçümünün diğer ölçüm metodları ile karşılaştırılması
Comparison of temporal artery thermometer and other body temperature measurement methods in children aged 1-36 months
CANSU AYDIN
Tıpta Uzmanlık
Türkçe
2022
Çocuk Sağlığı ve HastalıklarıAnkara Yıldırım Beyazıt ÜniversitesiÇocuk Sağlığı ve Hastalıkları Ana Bilim Dalı
DOÇ. DR. GÜLSÜM İCLAL BAYHAN
- Pediatrik malignitelerde febril nötropeni ataklarının değerlendirilmesi
Evaluation of febrile neutropenia episodes in pediatric malignancies
ZEYNEP GÜNAL TÜRK
Tıpta Uzmanlık
Türkçe
2020
Çocuk Sağlığı ve HastalıklarıSağlık Bilimleri ÜniversitesiÇocuk Sağlığı ve Hastalıkları Ana Bilim Dalı
DOÇ. DR. CENGİZ BAYRAM
- Ankara üniversitesi Tıp Fakültesi çocuk onkoloji servisinde 2008-2023 yılları arasında takip edilen febril nötropeni tanılı hastaların klinik, enfeksiyöz ve izlem özelliklerinin değerlendirilmesi
Evaluation of the clinical, infectious, and follow-up characteristics of patients diagnosed with febrile neutropenia monitored between 2008 and 2023 in the pediatric oncology unit of Ankara University Faculty of Medicine
DUYGU EMER DİLAVER
Tıpta Uzmanlık
Türkçe
2025
Çocuk Sağlığı ve HastalıklarıAnkara ÜniversitesiÇocuk Sağlığı ve Hastalıkları Ana Bilim Dalı
PROF. DR. HANDAN UĞUR DİNÇASLAN
- Nötropenik COVID-19 hastalarında klinik özellikler ve prognozun araştırılması
Başlık çevirisi yok
REZAN AKTERT
Tıpta Uzmanlık
Türkçe
2023
Acil TıpSağlık Bilimleri ÜniversitesiAcil Tıp Ana Bilim Dalı
PROF. DR. ÖZGÜR DİKME
- Febril nötropenik hastalarda kullanılan antibiyotiklere bağlı gelişen yan etkilerin karşılaştırılması
Comparison of adverse effects associated with antibiotics use in febrile neutropenic patients
HAFİZE KURT
Tıpta Uzmanlık
Türkçe
2012
HematolojiCelal Bayar Üniversitesiİç Hastalıkları Ana Bilim Dalı
PROF. DR. ÜLKÜ ERGENE