Küçük alanlarda farklı foton enerjilerine ait output faktörlerinin çeşitli dedektörler kullanılarak karşılaştırılması
Comparison of output factors of different photon energy which using various detectors in small fields
- Tez No: 526333
- Danışmanlar: PROF. DR. DENİZ YALMAN
- Tez Türü: Yüksek Lisans
- Konular: Fizik ve Fizik Mühendisliği, Physics and Physics Engineering
- Anahtar Kelimeler: Radyasyon Dedektörleri, Küçük Alan Dozimetrisi, Output Faktörü, Dedektörler, Foton, Radyasyon, Radyasyon dozajı, Radyasyon etkileri, Radyometri, Radyoterapi, Çıkış sinyali, Radiation Detectors, Small Field Dosimetry, Output Factor, Detectors, Photon, Radiation, Radiation dosage, Radiation effects, Radiometry, Radiotherapy, Output signal
- Yıl: 2018
- Dil: Türkçe
- Üniversite: Ege Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyasyon Onkolojisi Ana Bilim Dalı
- Bilim Dalı: Sağlık Fiziği Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
Yoğunluk ayarlı radyoterapi ve stereotaktik tedavi teknikleri gibi gelişmiş radyoterapi uygulamalarında sıklıkla kullanılan küçük radyasyon alanları, genel olarak 4 cm × 4 cm'den küçük boyutlu alanlar olarak kabul edilir. Küçük alanların dozimetrisi; lateral elektronik denge kaybı, dik doz değişimleri ve hacim ortalama etkisi gibi nedenlerden dolayı büyük alanlarınkine göre daha hassas ve karmaşıktır. Bunlardan lateral elektronik denge kaybı ve dik doz değişimleri ışınla ilgili problemlerdir; hacim ortalama etkisi ise dedektörle ilgilidir. Bu çalışmanın amacı, klinikte mevcut bazı dedektörler kullanılarak küçük alanlarda output ölçümleri için bu dedektörlerden hangilerinin daha uygun olduğunu belirlemektir. Çalışmada TLD ve farklı hacimlerde beş adet iyon odası kullanılmış olup, referans derinlikte (d=10 cm) ve farklı alan boyutları için 6 MV ve 18 MV foton enerjilerinde output ölçümleri alınmıştır. Bu ölçümlerden elde edilen veriler 10 cm × 10 cm referans alan boyutunun output değerine normalize edilerek output faktörü (OF) değerleri bulunmuştur. Output faktörlerinin istatistiksel analizi SPSS programında, Friedman testi uygulanarak yapılmıştır. Her iki enerji için dedektörleri 6'lı ve 4'lü gruplar halinde karşılaştırdığımızda çok küçük ve konvansiyonel küçük alanların OF değerleri arasında anlamlı farkın olduğu saptanmıştır. Büyük alanlarda (≥4×4 cm2) dedektörlere ait OF değerlerinin standart sapması 6 MV için %3'ten küçük, 18 MV için de %2'den küçüktür. Alan boyutu küçüldükçe (<4×4 cm2) dedektörlerin OF değerleri arasındaki fark giderek artmaktadır. Özellikle çok küçük alanlarda (≤1,5×1,5 cm2) standart sapmalar 6 MV için %10'a, 18 MV için %9'a kadar artmıştır. Mini dedektörler (CC01 ve CC04) mikro dedektörlere (TLD) göre daha büyük hacimli olmasına rağmen hava eşdeğeri yoğunlukta duvar materyaline sahip oldukları için çok küçük alanlardaki OF değerleri her iki enerjide de daha büyüktür. Böylece küçük alan OF ölçümleri için kullanılacak en ideal dedektörün“en küçük hacimli”dedektör olduğunun söylenemeyeceği ve dedektör seçiminde dedektörlerin etkin ölçüm hacminin yanı sıra yapı materyalinin özelliklerinin de göz önünde bulundurulmasının gerekli olduğu belirlenmiştir.
Özet (Çeviri)
Small radiation fields which commonly used in advanced radiotherapy applications, such as intensity modulated radiotherapy and stereotactic treatment techniques, are generally considered to be fields smaller than 4 cm × 4 cm. Small field dosimetry is more sensitive and complex than large fields due to lack of lateral electronic equilibrium, steep dose gradients and volume averaging effect. Lack of lateral electronic equilibrium and steep dose gradients are radiation related problems; the volume averaging effect is related to the detector. The aim of this study is to determine which of these detectors are more suitable for output measurements in small fields using some clinically available detectors. In the study TLD and five ionization chambers in different volumes were used and output measurements were taken at reference depth (d = 10 cm) and 6 MV and 18 MV photon energies for different field sizes. The data obtained from these measurements were normalized to the output value of the 10 cm × 10 cm reference field size and the output factors (OF) were found. Statistical analysis of the output factors was performed by using the Friedman test in the SPSS program. When we compare detectors in groups of 6 and 4 for both energies, it was determined that there is a significant difference between OF of very small and conventional small fields. Standard deviations of OF for detectors in large fields (≥4 × 4 cm2) are smaller than 3% for 6 MV and less than 2% for 18 MV. As the field size decreases (<4 × 4 cm2), the difference between the OF of detectors is increasing. Especially in very small fields (≤1,5 × 1,5 cm2) standard deviations increased to 10% for 6 MV and to 9% for 18 MV. Although the mini detectors (CC01 and CC04) have a larger volume than the microdetectors (TLD), the OF in very small fields are larger on both energies since they have wall material with air equivalent density. Thus, it is determined that the ideal detector to be used for small fields OF measurements can not be said to be the“smallest volume”detector, and that it is necessary to consider the properties of the building material as well as to the effective measurement volume of the detectors in the detector selection.
Benzer Tezler
- Küçük alanlara ait dozimetrik parametrelerin farklı detektörlerle ölçülmesi
Measurement of parameters for small field dosimetry with various detectors
SEVTAP YILDIRIM
Yüksek Lisans
Türkçe
2018
Fizik ve Fizik Mühendisliğiİstanbul Medipol ÜniversitesiSağlık Fiziği Ana Bilim Dalı
DOÇ. DR. HİLAL ACAR DEMİR
- Muon pair production at ultra-relativistic energies
Yüksek-rölativistik enerjilerde müon çifti üretimi
SEVGİ KARADAĞ
Yüksek Lisans
İngilizce
2018
Fizik ve Fizik Mühendisliğiİstanbul Teknik ÜniversitesiFizik Mühendisliği Ana Bilim Dalı
PROF. DR. MEHMET CEM GÜÇLÜ
- 6MV ve 18 MV foton enerjilerinde çok yapraklı kolimatör ve seroband alaşım blokların geometrik ve dozimetrik olarak karşılaştırılması
Dosimetric and geometric comparasion of multi leaf collimators and cerroband alloys at 6 and 18 MV photon energies
GÖRKEM GÜNGÖR
Yüksek Lisans
Türkçe
2007
OnkolojiHacettepe ÜniversitesiKlinik Onkoloji Ana Bilim Dalı
PROF. DR. MURAT GÜRKAYNAK
- Searches for supersymmetry with deep learning
Derin öğrenim metodlarıyla süpersimetrinin araştırılması
İLKNUR KÖSEOĞLU
Yüksek Lisans
İngilizce
2016
Fizik ve Fizik Mühendisliğiİstanbul Teknik ÜniversitesiFizik Mühendisliği Ana Bilim Dalı
DOÇ. DR. MUAMMER ALTAN ÇAKIR
- In0,53Ga0,47As/In0,52Al0,48As heteroeklemlerin elektronik transport özellikleri ve iki-boyutlu elektronların güç kaybı mekanizmalarının Shubnikov -de Haas osilasyonları yöntemiyle araştırılması
Investigation of the electronic transport properties and power loss mechanisms of the two-dimensional electrons in In0,53Ga0,47As/In0,52Al0,48As heterojunctions by Shubnikov -de Haas oscillations technique
ENGİN TIRAŞ
Doktora
Türkçe
2002
Fizik ve Fizik MühendisliğiHacettepe ÜniversitesiFizik Mühendisliği Ana Bilim Dalı
PROF. DR. MEHMET CANKURTARAN