Radyoterapide kullanılan tedavi planlama sistemi için bağımsız“monitor unit”doğrulaması
Independent monitor unit verification for treatment planning system used in radiotherapy
- Tez No: 395415
- Danışmanlar: YRD. DOÇ. DR. NINA TUNÇEL
- Tez Türü: Yüksek Lisans
- Konular: Onkoloji, Fizik ve Fizik Mühendisliği, Oncology, Physics and Physics Engineering
- Anahtar Kelimeler: Belirtilmemiş.
- Yıl: 2015
- Dil: Türkçe
- Üniversite: Akdeniz Üniversitesi
- Enstitü: Fen Bilimleri Enstitüsü
- Ana Bilim Dalı: Fizik Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Hasta bazlı kalite kontrol yöntemlerinden biri bağımsız MU doğrulamasıdır. Akdeniz Üniversitesi Fen Fakültesi Fizik ABD'ında yapılan bu çalışmada CMS XIO tedavi planlama sisteminde bulunan“XIO Standalone”bağımsız MU doğrulama yazılımının kabul testleri yapılarak klinikte hasta bazlı kalite kontrol için kullanılması amaçlandı. Kabul testlerinde kare açık, asimetrik açık ve kama filtreli kare alanlar için TPS'ten hesaplanan doz, IBA marka blue phantom2 su fantomunda ölçülüp“XIO Standalone”bağımsız MU doğrulama yazılımı ile hesaplandı.“XIO Standalone”kullanılarak hesaplanan dozun ölçüm ve TPS ile uyumlu olduğu görüldü. Daha sonra IBA I'mrt fantomunda 6 ve 15 MV foton enerjili altı alanlı konformal plan yapıldı. Bu planlar için“XIO Standalone”kullanılarak demet başına doz hesaplandı. Planın her demeti için I'mRT fantom ve FC65-P iyon odası kullanılarak ölçüm yapıldı. Tedavi masasının altında kalan demetlerde“XIO Standalone”ile TPS'te hesaplanan dozların ölçümden elde edilen dozlara göre farkları yüksek bulundu. Bu farkı ortadan kaldırmak için CMS Monaco konturlama sisteminde bulunan iBeam evo hazır masa modeli plana eklendi. Masa bileşenlerinin elektron yoğunluk değerleri sisteme girildi. XIO TPS'in masayı hesaplamaya katması için masa dış kontura dahil edildi. Aynı MU'lar ile TPS'te yeniden hesaplama yapıldı. Daha sonra“XIO Standalone”ile MU doğrulaması yapıldı. Masa eklendikten sonra TPS'te hesaplanan dozlar ile ölçümden elde edilen dozlar uyumlu bulundu. Ancak“XIO Standalone”yazılımının hesaplama ortamını su olarak kabul etmesi nedeniyle MU doğrulamasının ölçüme göre farkı tolerans dışında bulundu. Bu sorun için çözüm önerisi üretildi ve klinik kullanım için aksiyon seviyeleri belirlendi.
Özet (Çeviri)
Independent MU verification is one of the methods for patient based quality control. This study was carried out at the Akdeniz University Faculty of Science Department of Physics and was aimed to perform acceptance and commissioning“XIO Standalone”independent MU verification software in CMS XIO TPS. For acceptance tests the square open fields, asymmetric open fields and wedged filter fields was created in the TPS. Than the calculated dose from“XIO Standalone”was compared with obtained dose from TPS and measured dose by IBA blue phantom2 water phantom. The calculated dose from“XIO Standalone”was comparable with measured dose and calculated dose from TPS. Later, six-field conformal plans were generated for 6 and 15 MV photon beams for IBA I'mRT phantom. The dose was calculated for each beam in“XIO Standalone”. The dose of each beam was measured by FC65-P ion chamber and I'mRT phantom. The calculated dose from“XIO Standalone”didn't accordance with measured dose and calculated dose from TPS for beams that located under the treatment table. The iBeam Evo ready model table top was added in a created plan to eliminate this difference. The electron density value of table top compositions was entered into the system. Table top was included to outer contour for calculation in XIO. Recalculation was made with same MU in TPS. Then MU verification was made in“XIO Standalone”again. The measured doses and calculated doses from TPS were consistent after the table was added in the system. However,“XIO Standalone”according to measurement dose difference increased because Standalone hasn't inhomogeneity correction. The solution to this problem has been produced and the action levels were determined for clinical use.
Benzer Tezler
- Akciğer yerleşimli tümörlerde farklı radyoterapi teknikleri ve tedavi planlama algoritmalarının hedef tümör hacmi ve kritik organlar üzerindeki etkisinin değerlendirilmesi
Evaluation of different radiotherapy treatment techniques and treatment planning algorithms on target tumor volume and critical organs for lung tumors
AYÇA ÇAĞLAN
Doktora
Türkçe
2024
Radyasyon OnkolojisiSağlık Bilimleri ÜniversitesiRadyasyon Biyolojisi Ana Bilim Dalı
PROF. DR. BAHAR DİRİCAN
- Bağırsak hasar modellerinde kemik iliği mezenkimal kök hücre ve bağırsak mezenkimal kök hücrelerin rejeneratif ve immün modülatuvar etkilerinin karşılaştırılması
Comparison of regenerative and immune modulatory effecets of bone marrow and intestinal mesenchymal stem cells i intestinal damage models
BURCU PERVİN
Doktora
İngilizce
2025
BiyolojiHacettepe ÜniversitesiKök Hücre Ana Bilim Dalı
DOÇ. DR. FATİMA AERTS KAYA
- Özofagus kanseri radyoterapisinde hasta solunum hareketinin doz dağılımına etkisinin farklı planlama teknikleri ile araştırılması
Investigation of the effect of patient respiratory action on dose distribution in esophageal cancer radiotherapy with different planning techniques
MÜGE ÖZKIRIM
Yüksek Lisans
Türkçe
2020
Radyasyon Onkolojisiİstanbul ÜniversitesiTemel Onkoloji Ana Bilim Dalı
DR. ÖĞR. ÜYESİ NAZMİYE DÖNMEZ KESEN
- Radyoterapide kullanılan üç boyutlu tedavi planlama sistemi için doğrusal hızlandırıcı tedavi cihazından elde edilen elektron ölçüm sonuçlarının incelenmesi
Investigation of electron measurement results obtained from lineer accelerators for three dimension planning system at radiotherapy
EBRAR İNAÇ
- Radyoterapide kullanılan iki farklı tedavi planlama sistemi ile yüksek gradlı beyin tümörlü hastaların tedavi planlarının karşılaştırılması
Comparison of two different treatment planning systems used in radiotherapy and treatment plans of patients with high grade brain tumor
ELİF YEŞİL
Yüksek Lisans
Türkçe
2021
Radyasyon OnkolojisiNecmettin Erbakan ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
DOÇ. DR. MERYEM AKTAN