Yoğunluk ayarlı radyoterapi tekniğinin uygulanmasında kullanılan küçük alanların karakteristiklerinin incelenmesi

Examination of the characteristics of small fields used in intensity modulated radiotherapy technique

  1. Tez No: 373936
  2. Yazar: HÜLYA ÖZDEMİR
  3. Danışmanlar: YRD. DOÇ. DR. NINA TUNÇEL
  4. Tez Türü: Yüksek Lisans
  5. Konular: Onkoloji, Radyoloji ve Nükleer Tıp, Oncology, Radiology and Nuclear Medicine
  6. Anahtar Kelimeler: Belirtilmemiş.
  7. Yıl: 2014
  8. Dil: Türkçe
  9. Üniversite: Akdeniz Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Fizik Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Bu çalışma, Elekta marka Synergy lineer hızlandırıcı cihazında üretilen fotonların 6 MV demet enerjisi kullanıldı. Çalışmada yoğunluk ayarlı radyoterapi tekniğinde kullanılan küçük alanların karakteristiklerini incelemek için rölatif ve mutlak doz ölçümleri yapıldı ve birbirleriyle karşılaştırıldı. Çalışma kapsamındaki alanlar (1x1 cm2-10x10 cm2), cihazın merkezi ekseninde ve merkezi eksenin dışında üç farklı kolimatör ekseni üzerinde kaydırılarak yerleştirildi. Bu alanların yüzde derin dozunu (dozun derinliğe bağlı yüzde değişimi) ve profillerini belirlemek için, CC04 iyon odası ile Iba marka Blue su fantomu kullanıldı. Derin doz ve profillerden enerji tayini, dDmax derinliği, D20/D10 oranı, simetri, düzgünlük, penumbra ve FWHM verileri bulundu. Havada kolimatör saçılma faktörünü ölçmek için, CC04 iyon odası ile pirinç alaşımlı“build-up”başlık ve su fantomunun boş tankı kullanıldı. Doz verim ölçümleri için, katı su fantomda CC04 iyon odası ve termolüminesans dozimetri (TLD) kullanıldı. Ayrıca TLD ile 6 mm derinlikteki dozlar ölçüldü. Merkezi eksende yerleşmiş olan alanların küçülmesiyle dDmax derinliğinin ve D20/D10 oranının düştüğü görüldü. Demet simetrisinin tüm alanlarda limitler (±%3) içinde olduğu tespit edildi. Düzgünlük parametresinin 5x5 cm2'den küçük alanlarda %3 limitini aştığı bulundu. Bu nedenle düzgünlük parametresi yerine FWHM parametresinin kullanımının küçük alanlar için daha etkin olduğu tespit edildi. 10x10 cm2'nin kolimatör saçılma faktörü değerine göre, alan boyutunun küçülmesiyle kolimatör saçılma faktörü değerlerinin düştüğü görüldü. 16 mm derinlikteki doz verim değerleri alanın küçülmesiyle lineer bir azalım gösterdi. Aynı durum TLD ile ölçülen 6 mm derinlikteki (yüzey dozu) dozun değerlerinde de gözlendi. Alanların yerleşimleri merkezi eksenden uzaklaştıkça tüm kaydırmalarda dDmax derinliği yüzeye yaklaştı. X2, Y1 ve Diagonal düzlem kaydırmalarındaki her bir alanın profillerinden bulunan D20/D10 oranının düştüğü görüldü. Alanların merkezi eksenden kaymasına göre kolimatör saçılma faktörünün değeri %3-%6 oranında arttı. İyon odası ve TLD ile elde edilen doz verim sonuçlarında da bu artış oranı görüldü. Ayrıca, 2x2 cm2 alan boyutuna kadar bu dedektörlerin sonuçları birbirleriyle uyumluluk gösterdi. Genel olarak, alanların yüzey dozu tüm alan kaydırmalarında arttı. Konvansiyonel tekniklerle kıyaslandığında, YART tekniğinin dozimetrisi alan boyutlarının çok küçük olmasından dolayı önemlidir. Dolayısıyla, YART'ta doğru doz hesaplaması için küçük alanlar ve onların dozimetrik özellikleri TPS'de uygun bir şekilde modellenmeli ve sonrasında kalite kontrolü sağlanmalıdır. Tez kapsamındaki dozimetrik çalışmalar Akdeniz Üniversitesi Tıp Fakültesi Radyasyon Onkolojisi Anabilim Dalı'nda ve Denizli Devlet Hastanesi Radyoterapi Merkezi'nde yürütülmüştür.

Özet (Çeviri)

This study was performed by 6 MV photons produced by Elekta Synergy linear accelerator. The characteristics of the small fields, used in intensity modulated radiotherapy technique, were compared with each other in this study by relative and absolute dose measurements. Some of the fields in the scope of this study (1x1 cm2-10x10 cm2) were placed on central axis of the device and some of them were placed away from central axis of beam by shifting toward 3 different collimator axis. The IBA CC04 ion chamber and Blue water phantom were used to perform measurement of percentage depth dose (the percentage of dose depend on depths) and profiles on mentioned fields. The energy, dDmax, D20/D10, symmetry, flatness, penumbra and FWHM values were obtained from percentage depth dose and profiles. The CC04 ion chamber with brass alloy“build-up”cap and empty water phantom tank were used to define collimator scattering factor in the air. The output measurements of selected fields were performed by CC04 ion chamber and thermoluminescent dosimeter (TLD) in the solid water phantom. In addition, the 6 mm depth dose of each field was measured by TLDs. The dDmax depth and D20/D10 ratio were decreased depending on decreasing in size of fields that placed on the central axis. The beam symmetry value for all fields was found in the limit (+3%). The flatness parameter exceeded 3% limit value in fields smaller than 5x5 cm2. So, the FWHM was found to be much more effective parameter for small fields. According to the collimator scattering factor value of 10x10 cm2, the collimator scattering factor was decreased by decreasing the field size. Also, it was shown that the dDmax (16 mm) namely output values were linearly decreased by decreasing field size. The same behavior at the 6 mm (surface dose) was observed by TLD dose measurement values. When the field placements were moved away from central axis, the dDmax depth (16 mm) approached to the surface. Decreasing of D20/D10 ratio that calculated from profile of each field placed on the X2, Y1 and Diagonal directions, were seen. According to field displacement from central axis on all directions, the collimator scatter factor increased about 3% to 6%. The increment rate on output was detected in ion chamber and TLD results. Also, the detector results showed compatibility with each other up to 2x2 cm2. In general, the surface dose on each field increased by shifting in all direction. Compared to the conventional techniques, IMRT technique has more complicated dosimetry measurements because it employs smaller fields. Therefore, in order to calculate the correct dose in IMRT techniques the small fields and their dosimetric characteristics would be modeled appropriately in the treatment planning system and then the quality control should be provided. Dosimetric studies within the scope of the thesis were carried out at Akdeniz University Faculty of Medicine Department of Radiation Oncology and Denizli State Hospital Radiotherapy Centre.

Benzer Tezler

  1. Statik yoğunluk ayarlı radyoterapi tekniğinin iki farklı lineer hızlandırıcı ve tedavi planlama sisteminde dozimetrik olarak değerlendirilmesi

    Evaluation of Static Intensity Modulated Radiation Therapy as dosimetric in two different linear accelerators and treatment planning system

    FATİH KARAKÖSE

    Yüksek Lisans

    Türkçe

    Türkçe

    2013

    Fizik ve Fizik Mühendisliğiİstanbul Üniversitesi

    Temel Onkoloji Ana Bilim Dalı

    PROF. DR. HATİCE BİLGE

  2. Postoperatif endometrium kanserli hastalarda yoğunluk ayarlı radyoterapi ile rapidarc tekniğinin dozimetrik karşılaştırılması

    Dosimetric comparison of intensity modulated radiotherapy and rapidarc techniques for post-operative endometrial cancer

    OLGUN ELİÇİN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2013

    Kadın Hastalıkları ve Doğumİstanbul Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. GÜLYÜZ AKYIL ATKOVAR

  3. Meme kanserinde hibrit (konformal ve tanjansiyel yoğunluk ayarlı radyoterapi) planlama tekniğinin geleneksel alan içinde alan konformal radyoterapi tekniği ile karşılaştırılmasının retrospektif olarak incelenmesi

    Retrospective investogation of dosimetric comparasion of hybrid (conformal and tangential intensity modulated radiotherapy), field in field conformal radiotherapy techniques in breast cancer

    ECE AYFER CURA

    Yüksek Lisans

    Türkçe

    Türkçe

    2019

    OnkolojiBursa Uludağ Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. SİBEL ÇETİNTAŞ

  4. Erken evre glottik larenks radyoterapisinde yoğunluk ayarlı radyoterapi ile robotik kollu radyocerrahi tekniğinin dozimetrik karşılaştırılması

    Dosimetric comparison of intensity modulated arc therapy and robotic-ARM radiosurgery techniques in early stage glottic larynx radiotherapy

    SERDAR ALAY

    Yüksek Lisans

    Türkçe

    Türkçe

    2019

    Fizik ve Fizik Mühendisliğiİstanbul Üniversitesi

    Temel Onkoloji Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ MURAT OKUTAN

  5. Meme kanseri tedavisine üç boyutlu konformal izosentrik alan tekniği, alan içinde alan tekniği ve yoğunluk ayarlı radyoterapi tekniğinin dozimetrik karşılaştırılması

    Dosimetric comparison of three dimensional conformal isocentric technique, field in field technique and intensity modulated radiotherapy techniques for treatment of breast cancer

    ZELİHA DENİZ GÖBEL

    Yüksek Lisans

    Türkçe

    Türkçe

    2014

    Fizik ve Fizik MühendisliğiDokuz Eylül Üniversitesi

    Medikal Fizik Ana Bilim Dalı

    PROF. DR. İLKNUR BİLKAY GÖRKEN

    YRD. DOÇ. DR. HAKAN EPİK