Klorheksidin jelin farklı implant sistemlerinde abutment vidası üzerine etkilerinin karşılaştırmalı olarak incelenmesi
The effect of chlorhexidine gel on removal torque values of different dental implant systems
- Tez No: 342032
- Danışmanlar: PROF. DR. ENDER KAZAZOĞLU
- Tez Türü: Doktora
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Dental implant, Klorheksidin, Geri söküm torku iii, Dental implantasyon, Dental implantlar, Kemik vidaları, Dental Implants, Chlorhexidine, Removal Torque, Dental implantation, Dental implants, Bone screws
- Yıl: 2012
- Dil: Türkçe
- Üniversite: Yeditepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
mplant tedavisi gelişen teknolojisi sayesınde tek diş eksikliklerinde, pariyel ve total dişsizlik vakalarında sıklıkla ve başarı ile kullanılan vazgeçilmez bir tedavi seçeneğidir. İmplant tedavisi tüm bu avantajlarına rağmen mekanik ve biyolojik komplikasyonlarınedeniyle klinisyenleri sıkıntıya sokan durumlar oluşturmaktadır. Bu mekanik ve biyolojik komplikasyonlar içersinde en sık karşılaşılan komplikasyonlar vida gevşemesi ve kronik peri-implanitistir. Bu çalışmanın amacı implant abutment birleşimine lubrikant bir ajan olan klorheksidin jel uygulamasının abutment vidalarının geri sökülme tork değerleri üzerine biyolojik açıdan olduğu gibi mekanik bir katkısının da olup olmadığı incelemektir. Toplamda 24 adet yeni implant, abutment ve vidaları(Astra, Zimmer, Straumann Bone Level, Straumann Standart Plus) akrilik rezin içine gömüldükten sonra kuru ve klorheksidin uygulanacak iki gruba ayrıldı. Klorheksidin grubunda implant abutment vidasının arasındaki yivlere klorheksidin jel uygulanırken, kuru grupta herhangi bir ajan uygulanmayarak tüm abutmentlar dijital tork ölçer aleti ile firmaların önerdiği tork değerlerinde sıkılıp 5 dakika bekletildikten sonra söküldü. Bu işlem her implant için 15 kez tekrarlandı. Deneyin ikinci kısmında ise her iki grup termosiklus uygulamasından sonra elde edilen geri sökülme tork değerleri kaydedildi. Elde edilen tüm veriler student t test ve oneway anova testi ile istatistiksel olarak değerlendirildi (p<0.05). Strauman Bone Level implantları kuru ve klorheksidin uygulanmış haliyle başlangıç tork değerlerine göre en az kayıp ile geri sökülen implant olmuşlardır (2.66Ncm ve 3.91Ncm sırasıyla), Astra implantları ise kuru ve klorheksidinli ortamda en çok kayıp ile geri sökülen implant olmuştur (4.01Ncm ve 6.23Ncm sırasıyla). Sonuç olarak kuru ve klorheksidinli ortamda tekrarlayan sıkılma ve sökülme işlemini takiben başlangıç tork değerlerinde her zaman kayıp görülmüştür. Antimikrobiyal bir ajan olarak kullanılan klorheksidin jelin geri söküm tork değerlerini düşürdüğü, termosiklus uygulamasının ise herhangi bir etkisinin olmadığı görülmüştür.
Özet (Çeviri)
SUMMARY Ozdemir K. The Effect of Chlorhexidine Gel on Removal Torque Values of Different Dental Implant Systems. Yeditepe University Health Sciences Institute, PhD Thesis in Prosthetic Dentistry, İstanbul, 2012. Osseointegrated implants is a good represent now on acceptable treatment modality for complete or edentulous patients. Despite the recognized success, mechanical and biological failures associated with implant restorations represent a challenge to clinicians. Among the types of mechanical and biological failures abutment screw loosing and periimplantitis is still the most frequently reported complications. The aim of this study was to verify whether screw abutment lubrication with chlorhexidine gel can generate higher preload values compared to non-lubricated screws and determine if chlorhexidine has any extra mechanical advantage with it's biological advantage. Total of 24 new implants and abutments from (Astra, Zimmer, Straumann Bone Level, Straumann Standart Plus) each system were mounted in acrylic resin and the abutments were distrubuted in dry and chlorhexidine groups. In the chlorhexidine group the inner threads were filled with chlorhexidine gel. All abutments were fastaned with digital torque screw driver (DID-4) to their manifacturers recomended torque values and then detorqued after 5 minutes for 15 times. Thermocycling was performed for all implant systems both for dry and chlorhexidine groups for 1000 cycle. The mean detorque values were calculated and compared by student t-test and oneway anova test (p<0.05). Straumann Bone Level presented both in dry and chlorhexidine conditions showed the lowest loss of the initial torque values (2.66Ncm and 3.91Ncm respectively). Astra has the highest initial torque loss both for dry and chlorhexidine conditions (4.01Ncm and 6.23Ncm respectively). In conclusion there was always a loss in initial torque values when the removal torque was measured under both dry and chlorhexidine conditions. The chlorhexidine group presented in all implant abutment systems showed lower detorque values than the dry groups. Termocycling has no effect on removal torque values both in dry and chlorhexidine groups for all implant systems.
Benzer Tezler
- Implant bağlantı tipinin, klorheksidinin ve elastomerik tıkayıcıların dinamik koşullarda implant-abutment ara yüzündeki bakteriyel sızdırmazlığa etkisi
Effects of connection geometry and sealant agents on bacterial leakage along the implant-abutment interface:An in vitro study under loaded conditions
ARDA ÖZDİLER
Doktora
Türkçe
2017
Diş Hekimliğiİstanbul ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. FATMA GÜLBAHAR IŞIK ÖZKOL
- Dental implant içine yerleştirilen farklı konsantrasyonlarda klorheksidin ve polimer oranına sahip jellerin implant içi bakteriyel üreme üzerine kısa ve uzun dönem etkilerinin in vitro olarak incelenmesi
In vitro examination of short and long term effects of gels with varying clorhexidine to polymer concentration ratios installed into dental implants on bacterial growth within the implant
SEÇİL ÇOPUROĞLU
Doktora
Türkçe
2015
Diş HekimliğiHacettepe ÜniversitesiAğız, Diş ve Çene Cerrahisi Ana Bilim Dalı
DOÇ. DR. ALPER AKTAŞ
- Farklı irrigasyon maddelerinin kök kanal dolgu patlarının sızdırmazlığına etkisi
Sealing effectof different irrigation solutions upon root canal sealers
TÜLİN ERTAN
- Farklı iletken maddelerin elektrikli pulpa testinin cevabına etkisi
The effect of different conductive materials on response of electrical pulp test
EMİNE DEMİR ŞEREFLİ
Diş Hekimliği Uzmanlık
Türkçe
2019
Diş HekimliğiTokat Gaziosmanpaşa ÜniversitesiEndodonti Ana Bilim Dalı
DOÇ. DR. HÜDA MELİKE BAYRAM
- İn vitro olarak dentin bağlayıcı ajanların demiralize dentin üzerindeki antibakteriyel etkilerinin karşılaştırılması
Comparison of antibacterial effects of dentin bonding agents on demineralized dentin in vitro.
HANDE ŞAR SANCAKLI
Doktora
Türkçe
2007
Diş Hekimliğiİstanbul ÜniversitesiDiş Hastalıkları ve Tedavisi Ana Bilim Dalı
PROF. SAMİ BÜYÜKGÖKÇESU