Baş boyun fantom modelinde kontrast maddenin ve hava boşluklarının 3-boyutlu konformal radyoterapi planlamalarda doz dağılımına etkisi
The effects of contrast agent and air cavities on dose distribution of three-dimensional conformal radiotherapy plans in head and neck phantom model
- Tez No: 266987
- Danışmanlar: PROF. DR. FADİME AKMAN, YRD. DOÇ. DR. HAKAN EPİK
- Tez Türü: Yüksek Lisans
- Konular: Onkoloji, Fizik ve Fizik Mühendisliği, Oncology, Physics and Physics Engineering
- Anahtar Kelimeler: Kafa ve boyun neoplazmları, Kontrast maddeler, Neoplazmlar, Radyasyon, Radyasyon dozajı, Radyoterapi, Head and neck neoplasms, Contrast media, Neoplasms, Radiation, Radiation dosage, Radiotherapy
- Yıl: 2010
- Dil: Türkçe
- Üniversite: Dokuz Eylül Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Medikal Fizik Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Çalışmada baş boyun kanserleri radyoterapisinde, kontrast maddenin ve hava boşluklarının konvansiyonel ve konformal tedavi planlarındaki doz dağılımına etkisi karşılaştırılarak araştırıldı. Akrilik PMMA malzemesinden baş boyun bölgesine benzer fantom tasarlandı. Su ve kontrastlı suyun koyulabileceği, hava etkisinin incelenebileceği silindir bir kap yapıldı. Simülasyon için plaka, kavite, sulu ve kontrastlı su karışımının bulunduğu dört çekim görüntüsünden alınan ROI ile HU verilerinin ortalamaları HU< 125 yumuşak doku, silindir kap çeperi için ortalama HU>+125 kemik doku, su ve kontrastlı su ortalamaları sırasıyla 4.39 ile 27.18 HU olup sıvı ile uyumlu bulundu. Görüntülerde GTV, CTV ve PTV alanları konturlandı, her birinde konvansiyonel ve konformal olmak üzere 6 MV X ışını kullanılarak iki tedavi planı yapıldı. Toplam 200 cGy tek fraksiyonda, PTV'nin ortalama doz değerine ve %100'lük izodoza normalize edilerek hesaplandı. Plan sonuçları bağımsız gruplarda T testi ile hesaplanan p değeriyle analiz edildi. Plaka ve Kavite modellerinin konvansiyonel ve konformal plan karşılaştırmasında CTV dozu için p<0.05, boşluk plan lehine, anlamlı yüksek bulundu, boşluktan kaynaklanan saçılmaların etrafındaki CTV doz ortalamasında artışa neden olduğu düşünüldü. Plaka ve Su modellerinin konvansiyonel planında CTV, konformal planında GTV için su lehine anlamlı yüksek bulundu. Bu durum plaka, su ve silindir çeper arasındaki HU farklarından kaynaklanabilir, fakat günlük uygulamalarda anlamlı fark yaratacak düzeyde değildir. Su ile Kontrast'ın konvansiyonel ve konformal planlarında, GTV ile CTV' de anlamlı bir fark saptanmamıştır. Sonuçlara göre baş boyun bölgesinde planlama öncesi kontrast madde kullanımı tedavi dozunda artışa neden olurken, klinik açıdan etkisi ihmal edilebilir. Boşluklar geniş değilse, toplam dozda klinik açıdan ciddi değişimlere neden olmamakta, fakat boşluk duvarlarındaki normal doku doz birikmesi dikkate alınmalıdır.
Özet (Çeviri)
This comparative study was to investigate the radiotherapy of head and neck cancers by comparing the effects of the contrast agent and the air cavity on the patient dose in the conventional and conformal treatment plans. A phantom in the form of head/neck area has been designed with acrylic PMMA material. A cylindrical container was built in which the water and the contrast enhanced water could be placed in and the effect of air could be observed. For simulation the mean ROI and HU data gathered from the four images where slabs, cavity, water and contrast enhanced water were present was found to be HU<125, thus were compatible with soft tissue. HU was measured to be >125 for cylindrical container wall, thus compatible with the bone tissue. Water and contrast enhanced water are compatible with the liquid as they were measured to be 4.39 and 27.18 HU respectively. In the images GTV, CTV and PTV areas were countered and for each one of them two treatment plans that are conformal and conventional were planned with 6 MV X-ray. A single dose of 200 cGy were calculated to PTV and normalized to 100%. Plan results were analyzed with p-values of the t-test calculated from independent groups. In the comparison of slab and cavity models for the conventional and conformal plans the p-value of the CTV dose was found to be significantly high (<0.05) in favour of the cavity plan. The result was attributed to be a consequence of the scattering in the cavity which might have increased CTV mean dose. In the conventional plans of slab and water models CTV dose and in the conformal plan GTV dose were found to be significant in favour of water. This might be resulting from the HU differences in between the slab, water and the wall of cylindrical container, but for daily applications it is not significant enough to make any difference. In the conventional and conformal plans of water and contrast no significant difference was determined for GTV and CTV. According to the results the use of contrast agent before the treatment plan in the increase of the dose, whereas the clinical effect can be neglected. If the cavities are not wide, the total dose is not clinically affected in a serious manner, but the normal tissue dose accumulation in the cavity walls should be taken into account.
Benzer Tezler
- Baş-boyun kanseri tedavisinde yoğunluk ayarlı radyoterapi tekniğine canlı ortam doz ölçümünün katkısı
Contribution of in-vivo dose measurement on intensity modulated radiation therapy technique in head and neck cancer treatment
SEVDA SELEK
Yüksek Lisans
Türkçe
2013
Fizik ve Fizik MühendisliğiAkdeniz ÜniversitesiFizik Ana Bilim Dalı
YRD. DOÇ. DR. NINA TUNÇEL
- 3 boyutlu baskı teknolojisi kullanılarak hastaya özgün konturlarla oluşturulan fantom tasarımı ve üretimi ile mikro-çok yapraklı kolimatör tabanlı beyin ve başboyun yoğunluk ayarlı radyoterapi planlarının kalite kontrolü
Quality control of micro-multileaf collimator based brain and head-and-neck intensity modulated radiotherapy plans with patient-specific phantom design and production using 3D printing technology
SERCAN SUNEL
Yüksek Lisans
Türkçe
2017
Fiziksel Tıp ve RehabilitasyonHacettepe ÜniversitesiRadyoterapi Fiziği Bilim Dalı
PROF. DR. HÜSNÜ FADIL AKYOL
- Baş boyun kanserlerinin değişik tekniklerle (CD-60 ile) radyoterapisinde gözlerin almış olduğu dozların saptanması
The Evaluation of the radiation dose on the eyes during radiotherapy of head and neck cancers by different technics
ERCAN BALCI
Yüksek Lisans
Türkçe
1998
Göz HastalıklarıAtatürk ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
YRD. DOÇ. DR. B. DURSUN DEDE
- Baş-boyun IMRT'sinde yüzey dozlarının araştırılması
Investigation of surface doses in head and neck IMRT
ÖZLEM DORUK
Yüksek Lisans
Türkçe
2011
Radyoloji ve Nükleer Tıpİstanbul ÜniversitesiOnkoloji Ana Bilim Dalı
PROF. DR. GÖNÜL KEMİKLER
- Baş-boyun kanserlerinde giriş ve çıkış dozlarının in vivo dozimetri kullanılarak ölçülmesi ve tedavi planlama dozlarıyla karşılaştırılması
The measurement of entrance and exit doses with ın vivo dosimetry ın head and neck cancers and comparison with treatment planning doses
BURÇİN PAİDAR
Yüksek Lisans
Türkçe
2010
OnkolojiEge ÜniversitesiRadyasyon Onkolojisi Ana Bilim Dalı
PROF. DR. MUSTAFA ESASSOLAK