Titanyum dayanak tasarımının vida tutuculu implant restorasyonlarının retansiyonuna etkisi
Effect of titanium abutment design on retention of screw-retained implant restorations
- Tez No: 1013758
- Danışmanlar: PROF. DR. FATMA GÜLBAHAR IŞIK ÖZKOL
- Tez Türü: Doktora
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Titanyum dayanak, Zirkonya kuron, Retansiyon kuvveti, Dental implant, Rezin Siman, Titanium abutment, Zirconia crown, Retention force, Dental implant, Resin cement
- Yıl: 2026
- Dil: Türkçe
- Üniversite: İstanbul Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
- Bilim Dalı: Protetik Diş Tedavisi Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
Amaç: Bu çalışmanın amacı, farklı tasarımlara sahip Titanyum Base dayanaklarının (Ti-Base), simante edilmiş zirkonya kuronların retansiyonuna olan etkisinin incelenmesidir. Gereç ve Yöntem: Beş implant markasından (DT, Bİ, ST, SW, MO) farklı tasarımlara sahip on farklı Ti-Base grubu seçilmiş ve Ti-Baseler protetik yüksekliklerine göre iki gruba ayrılmıştır: <4,5 mm-K ve >4,5 mm-U. Her grup için on örnek kullanılmıştır. Toplamda 100 örnek test edilmiştir. Restorasyonlar aynı frezeleme ve sinterleme protokolleri kullanılarak üretilmiş ve standardize edilmiş koşullar altında aynı rezin siman ile simante edilmiştir. Tutuculuk kuvvetleri, restorasyonların Ti-Base'lerden ayrılması yoluyla özel olarak tasarlanmış bir kron çıkarma aparatı ve bir Universal test cihazı (Shimadzu Universal Testing Device) kullanılarak ölçülmüştür. Örnekler, dakikada 1 mm hızla başarısızlık (kopma) anına kadar çekme dayanımı testine tabi tutulmuştur. Veriler Jamovi istatistik yazılımı (Version 2.3, 2022) kullanılarak analiz edilmiştir (p < 0,05). Bulgular: Gruplar arasında çekme dayanımı bakımından anlamlı farklılık saptandı (p<0,001); en yüksek ortalama BİU ve MOU gruplarında (756 N), en düşük DTK grubunda (441 N) elde edildi. Tukey post-hoc analizinde gruplar dört homojen alt kümeye ayrıldı. Uzun Ti-Base'ler (681±155 N) kısa Ti-Base'lere (511±123 N) kıyasla anlamlı düzeyde yüksek retansiyon sergiledi (p<0,001). Uzunluk-bağımlı kazanım markaya göre belirgin biçimde farklılaştı; BİU (+%57) ve SWU'da (+%52) dramatik artış gözlenirken, MOU'da (+%11) sınırlı kazanım, STU'da ise anlamlı fark saptanmadı. Sonuç: Ti-Base tasarımı ve uzunluğu, zirkonya kuronların retansiyon kuvveti üzerinde anlamlı etkiye sahiptir. Uzun Ti-Baseler kısa Ti-Base'lere üstünlük sağlamış; uzunluktan elde edilen kazanım markaya göre belirgin biçimde farklılaşmıştır. Bu bulgular, üretici-bağımlı Ti-Base geometrisinin retansiyon performansında belirleyici rol oynadığını ortaya koymaktadır.
Özet (Çeviri)
Aim: The aim of this study is to investigate the effect of different Titanium Base abutment (Ti-Base) designs on loss of retention of zirconia crowns cemented to Ti-Bases. Materials and Methods: Ten Ti-Base groups with different designs from five implant brands were selected (DT,Bİ,ST,SW,MO) and Ti Base's were divided by prosthetic height (PH) into two groups; <4.5mm-K and 4.5>mm-U. Ten specimens were used for each group. Totally 100 specimens were tested. Restorations were fabricated using same milling, sintering protocols and cemented with the same resin cement under standardized conditions. Retention forces were measured by separating the restorations from the Ti-Base's using a custom crown removal device and a Universal testing machine (Shimadzu Universal Testing Device). The specimens tested with tensile strength test, 1mm per minute until failure. The data were analyzed using Jamovi statistical software (Version 2.3, 2022) (p < 0.05). Results: Significant differences in tensile strength were observed among the groups (p<0.001). The highest mean values were recorded in the BİU (gingival height [GH]: 1mm [PH]: 5.5mm and MOU (GH:1mm, PH 7mm) groups (756 N), whereas the lowest value was found in the DTK (GH:1mm, PH 4.5mm) group (441 N). Tukey post-hoc analysis categorized the groups into four homogeneous subsets. Long Ti-Base's (681±155 N) demonstrated significantly higher retention than short Ti Base's (511±123 N) (p<0.001). The magnitude of length-dependent improvement varied by brand. Conclusion: The height and design of Ti-Base abutments significantly influence zirconia crown retention, longer Ti-Bases showed superior performance, although the extent of improvement varied among manufacturers.
Benzer Tezler
- Farklı tekniklerle üretilmiş implant üstü dayanak ve kronların in-vitro olarak karşılaştırılması
In-vitro comparison of implant-supported single crowns and abutments manifactured with different techniques
İREM ÇÖTERT MERCAN
Doktora
Türkçe
2021
Diş HekimliğiEge ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. MÜBİN SITKI ULUSOY
- All-On-Four konsepti kullanılarak farklı tasarımlar ve materyallerle üretilen protezlerin stres dağılımının sonlu elemanlar analiz yöntemi ile incelenmesi
'evaluation of stress distribution of prosthesis produced with different designs and materials using the All-On-Four concept with finite element analysis method'
DAMLA GÜNEŞ ÜNLÜ
Doktora
Türkçe
2019
Diş HekimliğiErciyes ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
DOÇ. DR. KEREM KILIÇ
- Mandibulanın farklı ark formlarında dört implant ile tedavisinde implant açılarının ve trabeküler yüzey yapılı implantların kemik üzerine etkilerinin sonlu elemanlar analizi ile incelenmesi
Finite element analysis of the effects of implant angles and trabecular surface implants on bone in the treatment of mandible with four implants in different arch forms
AHMET İLTER ATAY
Doktora
Türkçe
2025
Diş Hekimliğiİstanbul ÜniversitesiOral İmplantoloji Ana Bilim Dalı
PROF. DR. SERDAR YALÇIN
- Titanyum abutment ve metal destekli restorasyonlara alternatif peek materyallerinin kullanımı sonrası peri-implant kemikte, implantta, kronda, abutmentta ve vidada oluşan streslerin sonlu elemanlar stres analizi ile karşılaştırması
Comparison the stresses occurring in the peri-implant bone, implant, superstructure, abutment and screw after using peek materials, which are alternative to titanium abutments and metal-supported restorations, by finite element stress analysis
SAMET TEKİN
Diş Hekimliği Uzmanlık
Türkçe
2017
Diş HekimliğiDicle ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
YRD. DOÇ. DR. YALÇIN DEĞER
- Farklı yüzey tasarımları ve titanyum içeriğine sahip dayanaklar ile implant üstü restorasyonlar arasındaki retansiyon kuvvet değerlerinin karşılaştırmalı olarak incelenmesi
Comparative study of alteration in retention force values of different titanium grades and abutment designs with implant-supported crowns
CANSU KARAHAN GÜNDÜZ
Diş Hekimliği Uzmanlık
Türkçe
2021
Diş HekimliğiSivas Cumhuriyet ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
DR. ÖĞR. ÜYESİ MELİH ÜLGEY