İmplant üretim toleransının implant-abutment arası mikro aralık ve sızdırmazlığa olan etkisinin değerlendirilmesi
Evaluation the effect of implant manufacturing tolerance on microgap and microleackage between implant-abutment
- Tez No: 828672
- Danışmanlar: PROF. DR. NEJAT NİZAM
- Tez Türü: Diş Hekimliği Uzmanlık
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Dental destekler, Dental sızma, Protezler ve implantlar, Üretim süreci, Dental abutments, Dental leakage, Prostheses and implants, Production process
- Yıl: 2022
- Dil: Türkçe
- Üniversite: Ege Üniversitesi
- Enstitü: Diş Hekimliği Fakültesi
- Ana Bilim Dalı: Periodontoloji Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Amaç: İmplant ve abutmentların üretim süreci belli bir tolerans aralığı içinde tamamlanır. İmplant ve abutment boyutlarında üretim toleransından kaynaklı farkların olması muhtemeldir. Bu çalışmanın amacı üretim toleransının, implant-abutment arası mikro aralık ve sızdırmazlığa olan etkisinin değerlendirilmesidir. Gereç ve Yöntem: İmplant-dayanak arasında mikro hareketin değerlendirilmesi için internal hibrit bağlantılı bir implant sisteminin iki farklı toleransa sahip implant ve abutmentları temin edildi. En düşük tolerans miktarına sahip örnekler Grup D, en yüksek tolerans miktarına sahip örnekler Grup Y olarak belirlendi. Her örnek iki yönlü bir çiğneme simülatörüne yerleştirildikten sonra, yük uygulandı ve bu sırada implant- abutment birleşimi üzerine devamlı olarak x-ışını verildi. X-ışını görünen ışığa dönüştürüldükten sonra yüksek hızda kayıt yapan bir cihaz ile video olarak kaydedildi. Elde edilen görüntülerin analiz edilmesi ile mikro hareket miktarları hesaplandı. Bulgular: Her iki grupta da implant-abutment arayüzünde 100N'a kadar olan yükler altında mikro aralık saptanmamıştır. 125N'dan itibaren ölçülebilen mikro aralık miktarı yüksek tolerans grubunda ortalama olarak daha fazla olsa da gruplar arasında istatistiksel anlamlı değildir. Her iki grup için de 125N, 150N, 175N ve 200N yük uygulamasıyla oluşan mikro aralık miktarları 25N, 50N ve 75N'a göre istatistiksel olarak anlamlı olacak şekilde artmıştır (hepsi için p<0.05). Sonuç: Hibrit bağlantılarda kuvvet artması ile mikro aralık görülebilmektedir. Üretim toleransı belirli bir seviyeye kadar mikro aralık miktarı üzerine etkili değildir.
Özet (Çeviri)
Aim: The manufacturing process of implants and abutments is achieved within a certain tolerance limit. There might be a difference between the dimensions of the implant and abutment. The aim of this study is to evaluate the effect of manufacturing tolerance on microgap formation between implant and abutment and sealing. Materials and Methods: Implants and abutments with an internal hybrid connection with two different tolerances were provided to evaluate the micromovement between the implant and abutment. The samples with the lowest tolerance amount were formed the Group D, and the samples with the highest tolerance amount were included in Group Y. After each specimen was placed in a bidirectional chewing simulator, the load was applied while an x-ray machined radiated continuously on the implant- abutment interface. After the x-ray was converted into visible light, it was recorded as video with a high-speed recording device. Micromovement amounts were calculated by analyzing the obtained images. Results: No microgap was detected at the implant-abutment interface under loads up to 100N in both groups. Although the amount of microgap that can be measured from 125N is higher in Group Y, the difference is not statistically significant between the groups (p>0.05). For both groups, the amount of microgap formed with the application of 125N, 150N, 175N and 200N loads increased statistically significantly. Conclusion: Micromovement can be observed parallel to the increasing force application in hybrid internal connections. The difference on tolerance may not have a significant affect at least up to a certain limit.
Benzer Tezler
- İmplant malzemelerinin Türkiye'de üretim perspektifi ve pazar analizi
Production and marketing perspectives of implant materials in Turkey
HÜSEYİN ERDEM
Yüksek Lisans
Türkçe
2005
Makine MühendisliğiAtatürk ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
PROF. DR. BÜNYAMİN AKSAKAL
PROF. DR. CAFER ÇELİK
- İmplant üstü dayanaklarda farklı koping üretim tekniklerinin kron tutuculuğuna ve marjinal uyuma etkisi: İn vitro çalışma
Effect of different coping production techniques on retention and marginal fit of implant supported crowns: An in vitro study
HAYRİYE ŞENTÜRK ÇEŞMECİ
Doktora
Türkçe
2015
Diş HekimliğiErciyes ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
DOÇ. DR. HASAN ÖNDER GÜMÜŞ
- İmplant destekli sabit protezlerde farklı üretim tekniklerinin implant-protez arasındaki uyuma etkisi
The effect of different manufacturing techniques on the fit of impant supported fixed prosthesis to implants
HÜSEYİN SULTAN
Doktora
Türkçe
2019
Diş Hekimliğiİstanbul ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. MEHMET CUDİ BALKAYA
- İmplant tasarımında eklemeli üretim kriterlerini dikkate alan bir topoloji optimizasyon yaklaşımının geliştirilmesi
Developing of a topology optimization approach considering additive manufacturing criteria in implant design
FAHRİ MURAT
Doktora
Türkçe
2024
Makine MühendisliğiErzurum Teknik ÜniversitesiMakine Mühendisliği Ana Bilim Dalı
PROF. DR. İRFAN KAYMAZ
- İmplant üstü metal seramik restorasyonlarda farklı altyapı üretim yöntemlerinin marjinal uyuma etkisinin değerlendirilmesi
Comparison of marginal fit of implant supported metal-ceramic restorations prepared by different techniques
BARIŞ PAŞALI
Diş Hekimliği Uzmanlık
Türkçe
2015
Diş HekimliğiOndokuz Mayıs ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. DUYGU SARAÇ