Engel geçiş katsayılarının yüksek enerjili X ışınları altında farklı zırhlama materyalleri için deneysel olarak belirlenmesi
Experimental determination of barrier transmission factors for different shielding materials unders high energy X-rays
- Tez No: 559933
- Danışmanlar: PROF. DR. AHMET BOZKURT
- Tez Türü: Yüksek Lisans
- Konular: Radyoloji ve Nükleer Tıp, Mühendislik Bilimleri, Fizik ve Fizik Mühendisliği, Radiology and Nuclear Medicine, Engineering Sciences, Physics and Physics Engineering
- Anahtar Kelimeler: Radyasyon ölçüm tekniği, Radyasyon-iyonizan, Radiation measurement technique, Radiation-ionizing
- Yıl: 2019
- Dil: Türkçe
- Üniversite: Akdeniz Üniversitesi
- Enstitü: Fen Bilimleri Enstitüsü
- Ana Bilim Dalı: Radyasyon Bilim ve Teknoloji Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
4, 6 ve 10 MV enerjilerde foton demeti üreten bir doğrusal hızlandırıcı cihazında birincil radyasyon zırhlama tasarımı amacıyla kurşun, çelik ve normal beton materyalleri için geniş demet engel geçiş katsayıları ölçümler yolu ile belirlenmiştir. Bu katsayıların elde edildiği ölçüm düzeneğinde kaynak bariyer arası mesafesi (SDD) 100 cm olarak ayarlanmıştır. Ölçümlerde dedektör olarak farmer tipi bir iyon odası kullanılmıştır. Ölçüm düzeneğinde dedektör, yeterli build-up sağlayacak 2-4 cm kalınlıkta su eşdeğeri katı fantom içerisine ve ışın merkezi ekseninde, engelin dış duvarından 30 cm uzağa yerleştirilmiştir. Çalışmada 3*3-10*10 cm2 aralığındaki farklı cihaz alan açıklıklarında beton materyali için engel geçiş katsayıları ölçümler yolu ile elde edilmiştir. Daha sonra bu değerler kullanılarak beton materyaline ait farklı cihaz alan açıklıklarında TVL değerleri hesaplanmıştır ve TVL değerlerinin cihaz alan açıklığına göre nasıl değiştiği incelenmiştir. Ayrıca çalışmada kaynak dedektör arası mesafenin 130, 230 ve 330 cm olarak ayarlandığı üç farklı ölçüm düzeneğinde her bir enerji için engelsiz bağıl doz değerleri ölçümler yolu ile elde edilmiştir ve bu değerlerin etkin alan boyutuna bağımlılığı da incelenmiştir. Bunun sonucunda geniş demet koşullarının 20*20 cm2 cihaz alan açıklığı sonrasında oluştuğu gözlemlenmiştir. Çalışmada kaynak dedektör arası 130 cm mesafede ve cihaz alan açıklığı 10*10 cm2'de kurşun, çelik ve beton materyalleri için elde edilen TVL değerleri, literatürde kullanılan değerlere uyarlanarak 40*40 cm2 cihaz alan açıklığı için TVL değerleri elde edilmiştir. Bu çalışmada belirlenen dar ve geniş demet TVL değerlerinin literatür değerleri ile uyumlu olduğu görülmüştür.
Özet (Çeviri)
Broad beam barrier transmission factors for lead, steel and ordinary concrete materials have been determined at 4, 6 and 10 MV photon energies in a linear accelerator device by means of measurements for design purposes of primary radiation shielding. In the setup where these factors were obtained, the distance from the source to barrier (SDD) was set to be 100 cm. A farmer type ionization chamber was used as the detector in the measurements. In the measurement set-up, the detector was placed 30 cm away from the outer wall of the barrier in a 2-4 cm thick solid phantom made of water equivalent material in the centre of the beam to provide adequate build-up. In the study, radiation transmission factors for concrete material were measured in field areas between 3*3-10*10 cm2. Then, using these values, TVL values of the corresponding field areas of concrete were calculated and the change of TVL values with respect to the field area TVL was evaluated. In addition, three different measuring setups where the distance from source to detector was varied as 130, 230 and 330 cm were studied for measuring barrier-free relative dose values at each beam energy and the dependence of these values on the effective field size were also examined. As a result, it has been observed that the broad beam conditions develop after 20*20 cm2 field area. In the study, TVL values were also obtained for lead, steel and concrete materials at a distance of 130 cm source to detector distance and at device field area of 10*10 cm2 and these values were adjusted to those in the literature for 40*40 cm2 to obtain TVL values of these materials for the maximum available field area. The narrow and broad beam TVL values determined in this study were found to be in good agreement with the literature.
Benzer Tezler
- Yeni Cami'nin akustik açıdan performans değerlendirmesi
Evaluation of the acoustical performance of the New Mosque
EVREN YILDIRIM
Yüksek Lisans
Türkçe
2003
Mimarlıkİstanbul Teknik ÜniversitesiMimarlık Ana Bilim Dalı
PROF. DR. SEVTAP YILMAZ DEMİRKALE
- Beta, gamma and neutron attenuation properties of PMMA/colemanite composites exposed to space radiation in low earth orbit
Alçak dünya yörüngesinde uzay radyasyonuna maruz kalan PMMA/kolemanit kompozitlerinin beta, gama ve nötron zayıflatma özellikleri
MEHMET ORHAN ŞİRİN
Yüksek Lisans
İngilizce
2024
Nükleer Mühendislikİstanbul Teknik ÜniversitesiNükleer Araştırmalar Ana Bilim Dalı
PROF. DR. NİLGÜN BAYDOĞAN
- Monitoring tropospheric water vapor variations with ppp during severe weather
Troposferik su buharı değişimlerinin ppp ile kötü hava koşullarında izlenmesi
ENGİN TUNALI
Doktora
İngilizce
2018
Mühendislik Bilimleriİstanbul Teknik ÜniversitesiGeomatik Mühendisliği Ana Bilim Dalı
DOÇ. DR. MUSTAFA TEVFİK ÖZLÜDEMİR
- Planning and stochastic evaluation of combined cooling heat and power systems under uncertainty
Belirsizlik durumlarında trijenerasyon sistemlerinin planlanması ve stokastik değerlendirilmesi
İBRAHİM ERSÖZ
Doktora
İngilizce
2019
Enerjiİstanbul Teknik ÜniversitesiEnerji Bilimi ve Teknolojileri Ana Bilim Dalı
PROF. DR. ÜNER ÇOLAK
- MODELLING AND ESTIMATION OF SHIP MOTIONS
GEMİ HAREKETLERİNİN MODELLENMESİ VE TAHMİNİ
ALPER ZİHNİOĞLU
Doktora
İngilizce
2023
Gemi Mühendisliğiİstanbul Teknik ÜniversitesiGemi İnşaatı ve Gemi Makineleri Mühendisliği Ana Bilim Dalı
DR. ÖĞR. ÜYESİ MELEK ERTOGAN