Farklı üst yapı materyalleri ile hazırlanan implant üstü sabit köprü protezlerinin implant, abutment, vida ve kemik üzerindeki stres dağılımının sonlu elemanlar analiz yöntemiyle incelenmesi
Investigation of implant, abutment, screw and bonestress distribution by using finite element method inimplant retained fixed bridge prosthesis prepared withdifferent superstructure materials.
- Tez No: 545977
- Danışmanlar: PROF. DR. KÖKSAL BEYDEMİR
- Tez Türü: Diş Hekimliği Uzmanlık
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: PEEK, monolitik zirkonya, zirkonya, sonlu elemanlar analizi, dental implant, Dental implantasyon, Dental implantlar, Dental materyaller, Dental protezler, Dental restorasyon, Dental stres analizi, Eterler, Kemik vidaları, Mini vidalar, PEEK, monolithic zirconia, zirconia, finite element analysis, dental implant, Dental implantation, Dental implants, Dental materials, Dental prosthesis, Dental restoration, Dental stress analysis, Ethers, Bone screws, Mini screws
- Yıl: 2018
- Dil: Türkçe
- Üniversite: Dicle Üniversitesi
- Enstitü: Diş Hekimliği Fakültesi
- Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Amaç: Farklı üst yapı malzemeleri kullanarak implant ve implant çevresi dokularda oluşan stresleri sonlu elemanlar stres analizi yöntemi ile değerlendirmektir. Gereç ve Yöntem: Çalışmamızda mandibula 2.premolar ve 2.molar bölgesine sırasıyla 3,3 mm ve 4,8 mm çapında titanyum implantlar yerleştirilmiştir. Sonlu elemanlar stres analizi yöntemi ile çalışma yapmak için kemik modellemesi yapılmıştır. Titanyum abutmentlar üzerine metal alt yapılı porselen, monolitik zirkonya, zirkonya altyapılı porselen ve PEEK materyalinden 4 farklı 3 üyeli köprü protezleri oluşturulmuştur. Okluzal yüzeye uygulanan dik ve oblik kuvvetler ile analizler gerçekleştirilmiştir. Bulgular: Oluşturulan 4 grupta da oblik kuvvetler dik kuvvetlerden daha yıkıcı etkiye sahip olduğu görülmüştür. Uygulanan kuvvetlerin etkisine bağlı olarak; kemikte oluşan stresler 4 grupta da sıkışma kuvvetleri etkisi altında ve implant, abutment ve vidada boyun bölgesinde, kronlarda ise konnektör ve kole bölgesinde bulunmuştur. Sonuç: Uygulanan kuvvetler ile streslerin genel olarak en az monolitik zirkonyada oluştuğu sonucuna varılmıştır.
Özet (Çeviri)
Aim: Aim of this study is to evaluate the stresses of the implant and implant surrounding tissues by using different superstructure materials by using the finite element method. Material and methods: In our study, titanium implants with a diameter of 3.3 mm and 4.8 mm were placed in the 2nd premolar and 2nd molar region of manadibula, respectively. Bone modeling was performed in order to analyse stress distribution by finite element method. 4 different 3-member bridge prostheses were designed by using porcelain fused metal, monolithic zirconia, porcelain fused zirconia and PEEK substructures on the titanium abutments. Analyzes were performed by vertical and oblique forces applied to the occlusal surface. Results: Oblique forces were found more destructive than vertical forces in all study grups. Stresses depending on the effect of the forces in the bone were affected by compression forces in 4 groups. Stresses in implant, abutment and screw were found around neck area. Stresses in crown were found around connector and margin areas. Conclusion: It is concluded that the stresses that occur as a result of the applied forces occur at least in zirconia generally.
Benzer Tezler
- İmplant üstü üç ve dört üyeli köprülerde metal üzerine preslenmiş porselen ile zirkonyum üzerine preslenmiş tam seramik ve konvansiyonel metal seramik restorasyonların marjinal ve internal uyum açısından karşılaştırılması
Comparison of marginal and internal fit of press-on-metal, all-ceramic zirconia systems and metal ceramics for three and four-unit implant-supported fixed partial dentures
EMRE BAYRAMOĞLU
Doktora
Türkçe
2014
Diş HekimliğiMarmara ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. YASEMİN KULAK ÖZKAN
- Abutment boyutu ile altyapı ve üstyapı kalınlıklarının posterior implant üstü zirkonyum kronların fraktür dayanımı üzerine etkisinin incelenmesi
Investigation the effect of abutment size and infrastructure-superstructure thickness on the fracture strength of posterior implant supporting zirconia crowns
AYŞE TUBA ÖĞRETEN
Doktora
Türkçe
2015
Diş Hekimliğiİnönü ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. MUSTAFA ÇETİN SUCA
YRD. DOÇ. DR. İSMAİL HAKKI UZUN
- Farklı implant üstü restoratif materyallerin implant kemik ara yüzeyine dinamik kuvvetler altında ilettiği streslerin incelenmesi
Investigation of stresses transmitted by different superstructure restorative materials to implant bone interface under dynamic loads
TOLGA AKOVA
Doktora
Türkçe
2002
Diş HekimliğiÇukurova ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
DOÇ. DR. HAKAN UYSAL
- Silikon esaslı maksillofasiyal protez malzemelerinin alt yapı malzemelerine bağlanma özelliklerinin araştırılması
The investigation of the bonding properties of silicone based maxillofacial prosthetic materials to the framework materials
BEGÜM YERCİ KOSOR
Doktora
Türkçe
2015
Diş HekimliğiEge ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. ATİLLA İSMAİL KESERCİOĞLU
- Titanyum alaşım materyalleri ile farklı bağlantı ajanları kullanılarak indirekt laboratuvar kompozit bağlantı dayanım değerlerinin incelenmesi
Evaluation of shear bond strengt between indirect laboratory composite and titanium surface while using different metal primers
NAZIM YÜKSEL YAZICIOĞLU
Doktora
Türkçe
2014
Diş HekimliğiYeditepe ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. ENDER KAZAZOĞLU