Farklı marka implant frezleri ile yapılan osteotomilerde kemikte meydana gelen sıcaklık değişimlerinin in-vitro olarak incelenmesi
In vitro evaluation of the temperature changes after bone osteotomy with different implant brand drills
- Tez No: 464945
- Danışmanlar: DOÇ. DR. AHMET MİHMANLI
- Tez Türü: Yüksek Lisans
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Dental implantasyon, Dental implantlar, Diş hekimleri, Diş hekimliği, Isı değişimi, Kemik ve kemikler, Osteotomi, Dental implantation, Dental implants, Dentists, Dentistry, Heat change, Bone and bones, Osteotomy
- Yıl: 2016
- Dil: Türkçe
- Üniversite: Bezm-İ Alem Vakıf Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Ağız, Diş ve Çene Cerrahisi Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Bu tez çalışmasında farklı firmalar tarafından üretilmiş implant frezleri ile yapılan drilleme esnasında, kemikte meydana gelen sıcaklık değişimlerinin in-vitro olarak incelenmesi amaçlanmıştır. Çalışmada taze sığır kaburgasından elde edilmiş kemik bloklar kullanılmıştır. Kemik bloklar üzerinde Ankylos, Astra Tech, Nobel Biocare, Bredent ve Straumann implant markaları tarafından üretilen frezler ile 2 kg sabit basınç altında 30±2 C'lik ortam sıcaklığında drilleme işlemi yapılmıştır. Kemik bloklara 5. ve 10. mm derinliklere iki adet K tipi termocouple sensör yerleştirilmiş ve implant frezine 1 mm mesafedeki sıcaklık değişimleri ölçülmüştür. Çalışmamızda, implant yuvası hazırlanması için farklı koşullarda çalışma modelleri oluşturulmuştur. Grup1'de ilk defa kullanılan frezlerle 150 rpm hızda irrigasyonsuz, grup 2'de ilk defa kullanılan frezlerle 1200 rpm hızda 40 ml/dk irrigasyon ile grup 3'de 30. defa kullanılan frezlerle 150 rpm hızda irrigasyonsuz, grup 4'de 30. Defa kullanılan frezlerle 1200 rpm hızda 40 ml/dk irrigasyon ile çalışılmıştır. Yapılan tüm osteotomi işlemleri 3,5±0,3 mm çaplı frezler ile 8 sn süre periyodunda gerçekleştirilmiş ve eşit zaman aralıklarında elde edilen sıcaklık değerleri kaydedilmiştir. Verilerin istatistiksel analizinde Repeated Measures ve Kruskall Wallis-H testleri kullanılmıştır. İmplant frezlerinin drilleme esnasında kemikte oluşturdukları sıcaklık değişiklikleri ile frezlerin arasında anlamlı bir fark gözlenmemiştir (p<0.05). Grupların hiçbirinde, çalışma boyunca kritik sıcaklık değerlerine (47C+) ulaşılmamıştır. 5. ve 10. mm derinliklerde kullanılan sensörlerdeki sıcaklık değişimlerinin de birbirine yakın olduğu görülmüştür. Ayrıca, irrigasyonsuz ve irrigasyonlu çalışılan sistemlerde ortalama sıcaklıkların birbirine yakın olmakla beraber; ancak başlangıç sıcaklığı çıkartılarak elde edilen fark değerlerinin, irrigasyonsuz sistemlerde anlamlı düzeyde yüksek olduğu hesaplanmıştır. Çalışmamızın sonuçları, implant frezlerinin kemik bloklarda üretici firmaların farklılığına (Ankylos, Astra Tech, Nobel Biocare, Bredent, Straumann firmaları) ve kullanım sayısına bağlı olarak önemli ölçüde sıcaklık artışlarına neden olmadığını göstermiştir. Ayrıca, irrigasyonlu ve irrigasyonsuz sistemlerin, önerilen koşullarda kullanıldığı sürece güvenli oldukları sonucuna ulaşılmıştır.
Özet (Çeviri)
The aim of this thesis was to investigate in vitro temperature changes during drilling in the bone by using implant drills which are produced by different companies. In this study, bone specimens obtained from fresh beef ribs were used. Drilling has been performed on the bone specimens by using the Ankylos, Astra Tech, Nobel Biocare, Bredent Straumann implant systems with 2 kilograms static downward force and 30±2 0C ambient temperature. Two K type thermocouples were placed in the 5 and 10 mm depth of bone specimens and temperature was measured at the distance of 1 mm from the implant drills. In our study, different experimental models were prepared to make implant site preparation. The experiment was based on four groups: in Group 1, drills were used for the first osteotomy at the speed of 150 rpm without irrigation; in Group 2 drills were used for the first osteotomy at the speed of 1200 rpm with irrigation (40 mL/min); in Group 3 drills were used for up to 30'th osteotomies at the speed of 150 rpm without irrigation and in Group 4 drills were used for up to 30'th osteotomies at the speed of 1200 rpm with irrigation (40 mL/min). Whole osteotomies were performed in eight minutes with 3.5±0.3 mm diameter drills and temperature values obtained in the eight equal time periods were recorded. Statistical analysis of the values was performed by using Repeated Measures and Kruskall Wallis tests. No significant difference was observed in each group for different implant brands according to temperature changes while drilling of the bone specimens (p<0.05). The critical temperature (47C+) was not recorded in any of the experimental models. In the 5 and 10 mm depth, the temperature change levels were similar to each other. Therefore, although the mean temperature values were similar in the irrigation and non-irrigation systems, the values obtained by subtraction of initial temperatures were significantly higher in non-irrigation groups. The results of this study showed that different implant drills (produced by Ankylos, Astra Tech, Nobel Biocare, Bredent, Straumann companies) and repeated drilling do not cause significant temperature rise during the osteotomy in bone specimens. Otherwise, we concluded that both irrigation and non-irrigation systems are safe in reasonable working conditions.
Benzer Tezler
- Aynı ticari marka implant frezleri ile yapılan osteotomilerde aynı frez kullanılarak“farklı hız, kuvvet ve kullanım sayısı”değişkenlerinin kemikte meydana gelen sıcaklık değişimleri üzerine etkilerinin in-vitro olarak incelenmesi
The in-vitro investigation of the effects of variables as“ different speed, force and number of uses ”on temperature changes in the bone using the same drills in the osteotomies produced with the same trademark implant mills
HARUN AKBÖRK
Diş Hekimliği Uzmanlık
Türkçe
2019
Diş HekimliğiGaziantep ÜniversitesiAğız, Diş ve Çene Cerrahisi Ana Bilim Dalı
DR. ÖĞR. ÜYESİ EBRU DENİZ KARSLI
- İki farklı soğutma sisteminin, farklı değişkenler eşliğinde implant frezlerinin kemik dokuda meydana getirdikleri sıcaklık üzerine etkilerinin in-vitro olarak incelenmesi
Investigation of the effects of two different irrigation systems accompanied by different variations on the heat caused by implant drills on bone an in-vitro study
EMRE TAŞDEMİR
Diş Hekimliği Uzmanlık
Türkçe
2021
Diş HekimliğiGaziantep ÜniversitesiAğız, Diş ve Çene Cerrahisi Ana Bilim Dalı
DR. ÖĞR. ÜYESİ EBRU DENİZ KARSLI
- Farklı marka implant abutmentlarının preparasyonu sırasında oluşan sıcaklığın değerlendirilmesi
Evaluation of temperature occurrence during preparation of different brand implant abutments
SELEN MERT
Doktora
Türkçe
2014
Diş HekimliğiYeditepe ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
PROF. DR. ENDER KAZAZOĞLU
- Koklear implant kullanıcılarında farklı konuşma işlemci modellerinin lokalizasyon ve gürültüde konuşmayı anlama becerileri üzerine etkisinin değerlendirilmesi
Assessment of the impact of different types of speech processors on localization and speech understanding in noise in cochlear implant users
EDA YALÇINKAYA
Yüksek Lisans
Türkçe
2022
Kulak Burun ve BoğazHacettepe ÜniversitesiOdyoloji Ana Bilim Dalı
DOÇ. DR. MERVE BATUK
- Posterior bölgede yeterli kemik yüksekliği bulunmayan hastalarda implant üstü sabit protez desteği için kısa boylu dental implantların kullanımının kemik kaybı açısından değerlendirilmesi
Evalution of marginal bone loss of short length dental implants used for implant supported fixed prosthesis at patients with insufficient bone height on posterior region
AYTAÇ YAŞAR
Doktora
Türkçe
2016
Diş HekimliğiErciyes ÜniversitesiProtetik Diş Tedavisi Ana Bilim Dalı
DOÇ. DR. AYŞEGÜL GÜLERYÜZ GÜRBULAK