Hacimsel ayarlı ark terapide lineer hızlandırıcının kalite kontrolü

Quality assurance of linear accelerator in volumetrik modulated ark therapy

  1. Tez No: 445232
  2. Yazar: ÜMMÜGÜL HASANOĞLU
  3. Danışmanlar: PROF. DR. ZEKİ GÖKAY KAYNAK
  4. Tez Türü: Yüksek Lisans
  5. Konular: Radyoloji ve Nükleer Tıp, Onkoloji, Radiology and Nuclear Medicine, Oncology
  6. Anahtar Kelimeler: Doz-cevap ilişkisi-ilaç, Kalite kontrol, Kolimatör sistemleri, Neoplazmlar, Parçacık hızlandırıcılar, Radyasyon, Radyometri, Radyoterapi, Dose-response relationship-drug, Quality control, Collimator systems, Neoplasms, Particle accelerators, Radiation, Radiometry, Radiotherapy
  7. Yıl: 2016
  8. Dil: Türkçe
  9. Üniversite: Uludağ Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Radyasyon Onkolojisi Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Hacimsel Ayarlı Ark Terapi, dinamik çok yapraklı kolimatör tekniklerini ve gantri arkını eş zamanlı olarak birlikte kullanır; bu nedenle kalite kontrolünün yapılması uygun bir tedavi için gereklidir. Bu çalışmada Uludağ Üniversitesi Radyasyon Onkolojisi Anabilim Dalın'da bulunan Elekta Synergy lineer hızlandırıcı cihazında 6 MV X-ışını kullanılarak Hacimsel Ayarlı Ark Terapi tedavi tekniğinde lineer hızlandırıcı cihazının kalite kontrolü gerçekleştirildi. Öncelikle lineer hızlandırıcının doz-monitor unit (MU) değerlerinin doğrusallığı, ışık alanı / radyasyon alanı doğruluğu, küçük MU değerleri için lineer hızlanıdırıcı performansı, Çok Yapraklı Kolimatör (ÇYK) pozisyon doğruluğu testi, ÇYK sızıntısı testleri incelendi. Farklı enerjiler için lineer hızlandırıcıda yüzde derin doz ( %DD), demet düzgünlüğü ve simetri ölçümleri yapıldı. Daha sonra çeşitli ÇYK kontrolü testleri yapıldı. Sonuç olarak lineer hızlandırıcı doğrusallık testine göre lineer hızlandırcı doz - MU değerleri doğrusaldır. Işık alanı / radyasyon alanı doğruluğu için ışınlanan film < 2 mm bulunmuştur. Küçük MU değerleri için lineer hızlandırıcı testinde 5 MU ve daha büyük değerler için sapma < % 2 ve 5 MU'dan küçük değerler için sapma < % 5 bulundu. ÇYK pozisyon doğruluğu < 1 mm ve ÇYK sızıntısı < % 5 bulundu. 6 MV ve 15 MV için cihazın % DD değerleri The British Journal of Radiology (BJR) Supplement 25 raporu ile uyumlu bulunmuştur. Demet düzgünlüğü ve simetri değerleri American Association of Physicist in Medicine (AAPM) Task Group 40 raporuna göre tolerans sınırları içerisinde bulunmuştur. ÇYK kontrolü testleri için lineer hızlandırıcıda ölçülen ile TPS'de hesaplanan dozlar karşılaştırılmış ve sonuçlar uyumlu bulunmuştur. Uygulanan teknik ve donanımla cihazın kalite kontrolü uluslararası protokollere uygun olarak gerçekleştirilmiş ve test edilen bu cihazın bu tedavi tekniğine uygun olduğu sonucuna varılmıştır.

Özet (Çeviri)

Volumetric modulated arc therapy (VMAT) involves the simultaneous use of dynamic multileaf collimator (DMLC) techniques and gantry arcing; appropriate quality assurance is therefor required. This study examines the quality assurance of a linear accelerator in volumetric modulated arc therapy technique by the Elekta Synergy linear Accelerator set up in Department of Oncology of Uludag Univercity at an energy level of 6 MV. First of all dose-monitor unit (MU) value linearity, accuracy of illuminated area / irradiated area, the linear accelerator performance for small MU delivery, test of Multileaf Collimator possition accuracy and interleaf leakage are considerated. % DD, flatness and symetry values in different energies are analysed for linear accelerator. Second, tests of Multileaf Collimators are analysed. Consequently, it is found that dose - MU value is linear. Accuracy of illuminated area / irrradiated area is < 2 mm. Linear accelarator performance for small MU values is < % 5 for < 5 MU and < % 2 for ≥ 5 MU. It is found that possition accuracy of MLC < 1 mm and İnterleaf leakage value is < % 5. % DD values of the device found in the tolerance zone according to The Bristish Journal of Radiology Supplement 25 Report for 6 MV and 15 MV. The dosimetric value of flatness and symetry found in the tolerance limits according to American Association of Physicist in Medicine (AAPM) Task Group Report 40. Measured and calculated doses are compared for MLC control tests and the results were consistent. The device quality assurance controls are made by the techniques and hardware conforming to international protocols and it's concluded that this device is suitable for this type of threatment.

Benzer Tezler

  1. İki boyutlu ve üç boyutlu görüntüleme teknikleri kullanılarak planlanan hedef oluşturma geometrik sınırlarının (marj) istatistiksel hesaplanması

    Statistical calculation of target creation geometric limits (marj) planned by using two-dimensional and three-dimensional imaging techniques

    RUMEYSA GEYİK

    Yüksek Lisans

    Türkçe

    Türkçe

    2018

    Fizik ve Fizik MühendisliğiAkdeniz Üniversitesi

    Fizik Ana Bilim Dalı

    DR. ÖĞR. ÜYESİ ŞERAFETTİN YALTKAYA

  2. Baş-boyun kanserlerinde yoğunluk ayarlı radyoterapi (YART)ve hacimsel ayarlı ark terapi (VMAT) tekniklerinin hastayaözel tedavi kalite güvencelerinin karşılaştırılması

    Comparison of intensity modulated radiationtherapy and volumetric modulated arc therapy methods of quality assuarence in head and neck cancers

    ABDURRAHMAN İSMAİL KORKMAZ

    Yüksek Lisans

    Türkçe

    Türkçe

    2015

    Radyasyon OnkolojisiUludağ Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. CANDAN DEMİRÖZ ABAKAY

  3. Baş boyun kanserli olgularda dört farklı hacimsel ayarlı ark terapi (VMAT) planının dozimetrik olarak incelenmesi: Retrospektif çalışma

    Examination of dosimetric study of four different volumetric modulated arc therapy (VMAT) plans in cases of head and neck cancer: Retrospective study

    OĞUZ AYDIN

    Yüksek Lisans

    Türkçe

    Türkçe

    2018

    Sağlık EğitimiBursa Uludağ Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    DOÇ. DR. CANDAN DEMİRÖZ ABAKAY

  4. Radyoterapi tedavisinde erken evre sol meme kanseri hastalarının nefes kontrollü ve kontrolsüz hacimsel ayarlı ark terapi tekniği ile kalp dozlarının karşılaştırılması

    Comparison of heart dosages with breath controlled and uncontrolled volumetric modulated arc therapy of early stage left breast cancer patients in radiotherapy

    EBRU ŞİMŞEK

    Yüksek Lisans

    Türkçe

    Türkçe

    2019

    BiyofizikMarmara Üniversitesi

    Biyofizik Ana Bilim Dalı

    DOÇ. DR. GÜNSELİ AYŞE GARİP

    DR. ÖĞR. ÜYESİ BORA TAŞ

  5. Baş-boyun kanserlerinin radyoterapisinde farklı vmat tedavi plan tasarımlarının plan kalitesi açısından karşılaştırılması

    Comparison of different vmat treatment plan designs in radiotherapy of head and neck cancers in terms of plan quality

    EZGİ YILMAZ

    Yüksek Lisans

    Türkçe

    Türkçe

    2021

    Radyasyon OnkolojisiEge Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. MUSTAFA ADNAN ESASSOLAK