Farklı yoğunluktaki kemiklerde mini-implant ring aparatının minivida stabilitesi üzerine etkisi
Efficacy of mini implant ring apparatus on the stability of mini screws in different bone densities
- Tez No: 340705
- Danışmanlar: DOÇ. DR. DERYA ÇAKAN
- Tez Türü: Doktora
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Minivida, Mini implant ring (MİR) aparatı, kortikal kemik yoğunluğu, primer stabilite, Dental implantasyon, Dental implantlar, Kemik dansitesi, Kemik ve kemikler, Kemik vidaları, Protezler ve implantlar, Mini screws, Mini implant ring (MIR) apparatus, cortical bone density, primary stability, Dental implantation, Dental implants, Bone density, Bone and bones, Bone screws, Prostheses and implants
- Yıl: 2013
- Dil: Türkçe
- Üniversite: Yeditepe Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Ortodonti Ana Bilim Dalı
- Bilim Dalı: Belirtilmemiş.
- Sayfa Sayısı: Belirtilmemiş.
Özet
Bu çalışmanın amacı, kortikal kemik yoğunluğunun minivida stabilitesine olan etkisini değerlendirmek ve mini implant ring (MİR) aparatının farklı yoğunluktaki kemiklerde minivida stabilitesi üzerine olan etkinliğini incelemektir. Bu amaçla çalışmada, 108 adet 1,6 mm çapında, 8 mm uzunluğunda konik minivida, 54 adet MİR aparatı ve farklı kortikal kemik yoğunluğundaki yapay kemikler kullanılmıştır. Kontrol gruplarını 20, 30 ve 40 pcf kortikal kemik yoğunluğuna sahip yapay kemiklere uygulanmış minividalar, test gruplarını ise yine 20, 30 ve 40 pcf yoğunluğuna sahip yapay kemiklere MİR ile uygulanmış minividalar oluşturmaktadır. Grupların maksimum uygulama torkları (MUT), kuvvet dirençleri (KD) ve maksimum çıkarma torkları (MÇT) ölçülmüştür. Veriler istatistiksel olarak değerlendirilmiştir. Farklı kortikal kemik yoğunluklarında ölçülen MUT, KD ve MÇT değerleri arasında istatistiksel olarak anlamlı fark bulunmuştur (p<0,001). Tüm kortikal kemik yoğunluklarında MİR grubunun, MUT, KD ve MÇT değerleri, kontrol grubunun MUT, KD ve MÇT değerlerine göre istatistiksel olarak anlamlı derecede yüksek bulunmuştur (p<0,001). Sonuç olarak, yapay kortikal kemik yoğunluğunun artması minividaların stabilitesini arttırmaktadır. Benzer şekilde MİR?le beraber uygulanan minividaların başlangıç stabiliteleri kortikal kemik yoğunlukları arttıkça artmaktadır. MİR aparatı uygulaması ankraj değerlerini olumlu yönde değiştirmektedir.
Özet (Çeviri)
The aim of this study is to evaluate the influence of cortical bone density on mini screw stability and to investigate the efficacy of mini implant ring apparatus (MIR) on the stability of mini screws in different bone densities. For this purpose, 108 conical mini screws (1.6 mm diameter, 8 mm length), 54 MIR apparatus and artifical bone samples with different cortical densities were used in this study. Control groups consisted of mini screws applied to artificial bone samples with densities of 20, 30 and 40 pcf whereas experiment group included MIR supplemented mini screws mounted to artificial bone samples with the same densities as the former group. Maximum insertion torque (MIT), force resistance (FR) and maximum removal torque values (MRT) were measured for each group. The findings were evaluated statistically. Statistically significant differences were found among MIT, FR and MRT values measured in different cortical bone densities (p<0.001). MIT, FR and MRT values for MIR group were found to be statistically significantly higher than those for control group in all cortical bone densities (p<0.001). To conclude, increase in cortical bone density also enhances the primary stability of mini screws. Likewise, stability of MIR supplemented mini screws increased accordingly with the bone densities. Application of MIR apparatus influences anchorage values in a positive manner
Benzer Tezler
- Kortikal kemik kalınlığı ve kortikal kemik yoğunluğundaki farklılıkların ortodontik mini-implantların primer stabilitesine olan etkisinin in-vitro olarak araştırılması
Evalution of the effects of differences in cortical bone thickness and cortical bone density on the primary stability of orthodontic mini-implants in-vitro.
EMİNE BUĞRA
- Farklı yoğunluktaki kemiklere yerleştirilen mini vidaların çapının ve tekrar kullanımının primer stabiliteye etkisinin İn-Vitro değerlendirilmesi
In Vitro evaluation of the impact of mini screw diameter and re-insertion on primary stability in bone with different densities
ELİF COŞKUNER UYSAL
Doktora
Türkçe
2025
Diş HekimliğiAltınbaş ÜniversitesiOrtodonti Ana Bilim Dalı
DR. ÖĞR. ÜYESİ PELİN ACAR ULUTAŞ
- Dental implant frezlerinin drillemesi sırasında farklı yoğunluktaki kemiklerde açığa çıkan ısının sonlu elemanlar analizi ile değerlendirilmesi
Evaluation of the heat released in bones of different density during the drilling of dental implant burs by finite element analysis
SELEN ONAT
Doktora
Türkçe
2023
Diş HekimliğiBaşkent ÜniversitesiAğız, Diş ve Çene Cerrahisi Ana Bilim Dalı
DOÇ. DR. NUR ALTIPARMAK
- Farklı implant uzunluklarının, tasarımlarının ve farklı kemik tiplerinin implant primer stabilitesi üzerine etkilerinin değerlendirilmesi: İn vitro bir çalışma
Evaluation of the effects of different implant lengths, designs, and bone types on primary implant stability: An in vitro study
BERFİN GÜRBÜZ
Diş Hekimliği Uzmanlık
Türkçe
2025
Diş HekimliğiAnkara Yıldırım Beyazıt ÜniversitesiPeriodontoloji Ana Bilim Dalı
DOÇ. DR. MAHMURE AYŞE TAYMAN
- Birinci metatarsofalanks ekleminin birleştirilmesinde (füzyonunda) stapler ve açı oluşumlarının deneysel ve sonlu elemanlar yöntemiyle incelenmesi
Experimental and finite element method investigation of stapler and angle formations in joining (fusion) of the first metatarsophalangx joint
GÜLİSTAN DURSUN
Yüksek Lisans
Türkçe
2023
Makine MühendisliğiSakarya ÜniversitesiBiyomedikal Mühendisliği Ana Bilim Dalı
DR. ÖĞR. ÜYESİ OSMAN İYİBİLGİN
DOÇ. DR. LEVENT BAYAM