CyberKnife robotik radyocerrahi cihazinda kullanilan küçük alanlarin dozimetrisinin farkli sistemler kullanilarak incelenmesi

The dosimetric evaluation of small fields of CyberKnife robotic radiosurgery unit with various dosimetric systems

  1. Tez No: 267429
  2. Yazar: TALİP YOLCU
  3. Danışmanlar: PROF. DR. MUSTAFA CENGİZ
  4. Tez Türü: Yüksek Lisans
  5. Konular: Onkoloji, Oncology
  6. Anahtar Kelimeler: Radyasyon, Radyasyon dedektörü, Radiation, Radiation detector
  7. Yıl: 2010
  8. Dil: Türkçe
  9. Üniversite: Hacettepe Üniversitesi
  10. Enstitü: Sağlık Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Klinik Onkoloji Ana Bilim Dalı
  12. Bilim Dalı: Radyoterapi Fiziği Bilim Dalı
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

Küçük alan dozimetrisi, yanal elektronik dengenin olmaması ve kullanılan ölçüm sisteminin boyutundan dolayı ciddi belirsizlikler içerir. Bu tez çalışmasında, farklı dedektörler ile CyberKnife robotik radyocerrahi cihazında kullanılan 5, 7.5, 10, 15, 20, 25, 30 ve 40 mm' lik kolimatörler için dozimetrik incelemeler yapıldı. Her bir kolimatör için farklı dedektörler ile verim faktörü eğrileri, yüzde derin doz eğrileri, 15, 50, 100 ve 200 mm derinliklerde profil eğrileri elde edildi. Her bir dedektörden elde edilen bu eğriler karşılaştırıldı. Diamond, 60008 ve 60012 diyot dedektörlerin verdiği cevaplar uyumlu bulunurken, 31010 ve 30013 iyon odalarının küçük alan dozimetrisi için uygun olmadığı saptandı. EBT film için penumra genişlemesi kabul edilebilir düzeyde iken YYG değerleri büyük çıktı. Pinpoint dedektör için YDD ve verim faktörü değerleri diamond dedektör ile uyumlu iken, profil değerleri kabul edilebilir düzeyde değildir ve literatür ile uyum göstermemektedir. Pinpoint dedektör için dikey konumda çözünürlük artmaktadır. Her bir dedektörün etkisini inceleyebilmek için elde edilen değerler tedavi planlama sistemine yüklenmeli ve aynı tedavi planlamaları yapılmalıdır. Gelecekte çok daha küçük hacimli ve doku eşdeğeri dedektörlerin üretilmesiyle bu belirsizlikler azalacaktır.

Özet (Çeviri)

Small field dosimetry implies serious uncertainities due to lateral electronic disequilibrium and dimensions of the detector used for measurements. In this thesis, dosimetric measurements were done with various detectors for the colimators having 5, 7.5, 10, 15, 20, 25, 30 and 40 mm diameters in CyberKnife robotic radiosurgery unit. For every individual collimator, using different detectors, output factor (OF) curves, percentage depth dose (PDD) curves and at the depths of 15, 50, 100 and 200 mm profile (off-center ratio, OCR) curves were obtained. The resulting curves of different detectors and also of diamond and pinpoint detectors for the positions paralel and vertical to the water surface were compared. The results obtained with diamond, 60008 and 60012 diode detectors were in aggrement but the results obtained with 31010 and 30013 ion chambers were not acceptable for small field dosimetry. For the EBT Gafchromic film altough penumbra broadening were acceptable, full width half maximum (FWHM) values were not. Percentage depth dose (PDD) and output factor curves of pinpoint detector were in aggrement with the diamond detector?s results but the profile results were not, and different from the literature. To make more accurate analyze, the data obtained with individual detectors should be loaded to the treatment planning system (TPS) and made the same treatment plans. In future, production of smaller and tissue equivalent detectors will give rise to more accurate dosimetric evaluations.

Benzer Tezler

  1. Cyberknife robotik radyocerrahi cihazında farklı anotomik yerleşimli tümörlerde gonadların aldığı dozun belirlenmesi

    Determination of gonad dose at cyberknife robotic radiosurgery for several tumor treatments

    GÖZDE DÜĞEL

    Yüksek Lisans

    Türkçe

    Türkçe

    2010

    OnkolojiHacettepe Üniversitesi

    Klinik Onkoloji Ana Bilim Dalı

    PROF. DR. FARUK ZORLU

  2. İleri evre küçük hücreli dışı akciğer kanserinde kemoterapi sonrası FDG-PET/BT'deki metabolik cevabın prognoz üzerine etkisi

    The prognostic relevance of response evaluation to chemotherapy using FDG-PET/CT in patients with advanced stage non-small-cell lung cancer

    ÇETİN ORDU

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2010

    OnkolojiYeditepe Üniversitesi

    İç Hastalıkları Ana Bilim Dalı

    DOÇ. DR. BAŞAK OYAN ULUÇ

  3. Robotik radyocerrahi uygulanan beyin ve spinal tümörlü hastalarda tedavi sonuçlarımızın değerlendirilmesi

    Therapeutic results in patients with brain and spinal tumors treated with robotic radiosurgery

    AYŞEN AYDIN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2011

    OnkolojiHacettepe Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. FARUK ZORLU

  4. Antioksidan ajanların deneysel olarak oluşturulan radyasyona bağlı akciğer fibrozu üzerinde etkinliğinin histopatolojik olarak değerlendirilmesi

    Histopathological evaluation of antioxidants' effectiveness on experimentally radiation induced lung fibrosis

    BİLGEHAN ŞAHİN

    Tıpta Uzmanlık

    Türkçe

    Türkçe

    2011

    Radyoloji ve Nükleer Tıpİstanbul Üniversitesi

    Radyasyon Onkolojisi Ana Bilim Dalı

    PROF. DR. ETHEM NEZİH ORAL

  5. CyberKnife robotik radyocerrahi cihazının tedavi planlama sisteminin dozimetrik kontrolü

    Dosimetric control of the treatment planning system of CyberKnife® robotic radiological surgery instrument.

    TAMER BAŞER

    Yüksek Lisans

    Türkçe

    Türkçe

    2011

    OnkolojiHacettepe Üniversitesi

    Klinik Onkoloji Ana Bilim Dalı

    PROF. DR. MURAT GÜRKAYNAK