Bütünüyle katı-hal kontakt amonyum-seçici PVC-membran kondüktometrik elektrotlar ve performans karakteristikleri

All solid-state contact ammonium-selective PVC-membrane conductometric electrodes and their performance characteristics

  1. Tez No: 135980
  2. Yazar: CELAL TURAK
  3. Danışmanlar: DOÇ. DR. İBRAHİM IŞILDAK
  4. Tez Türü: Yüksek Lisans
  5. Konular: Kimya, Chemistry
  6. Anahtar Kelimeler: PVC-membran Elektrotlar, Bütünüyle Katı-Hal, Seçici kondüktometri, Amonyum, Toprak Analizi, Amonyum bileşikleri, Elektrotlar, Kondüktometri, PVC-Membrane Electrodes, All Solid-State, Selective Conductometry, Ammonium, Soil Analysis, Ammonium compounds, Electrodes, Conductometry
  7. Yıl: 2003
  8. Dil: Türkçe
  9. Üniversite: Ondokuz Mayıs Üniversitesi
  10. Enstitü: Fen Bilimleri Enstitüsü
  11. Ana Bilim Dalı: Kimya Ana Bilim Dalı
  12. Bilim Dalı: Belirtilmemiş.
  13. Sayfa Sayısı: Belirtilmemiş.

Özet

11 BÜTÜNÜYLE KATI-HAL KONTAKT AMONYUM-SEÇİCİ PVC-MEMBRAN KONDÜKTOMETRİK ELEKTROTLAR VE PERFORMANS KARAKTERİSTİKLERİ ÖZET Nonaktin, dibenzo-18-crown-6 ve disüdo-18~crown-6 gibi crown-eter bileşikleri (nötral iyonoforlar olarak) kullanılarak tümüyle katı-hal kontakt PVC-membran NH/- seçici elektrotlar hazırlandı. Elektrotların kondüktometrik ve potansiyometrik performansları bilgisayarlı bir ölçüm sisteminde araştırıldı. Hazırlanan elektrotların tamamına yakını ana iyon çözeltisine karşı Nernst davranışı gösterirken (>60 mV/her 10 kat değişim), cevap zamanları kısa (<15 s) ve kullanım ömürleri uzundu (en az 1-2 ay). or -i Tüm elektrotlar 10“ -10”mol.L“ konsantrasyon aralığında ana iyon çözeltisine karşı doğrusal davranış sergilemekteydi. Elektrotların ana iyon yanında girişim yapan iyonlara karşı seçicilik katsayıları ayrı çözelti metodu ile hesaplandı. A ve D ile ifade edilen membran elektrotlar NHU+ iyonuna karşı Na+ ve K+ iyonları yanında oldukça yüksek kondüktometrik ve potansiyometrik seçicilik sergilediği, B ve C membran elektrotların ise çözeltide NH41”iyonuna karşı düşük düzeyde seçicilik sergilediği sonucuna varıldı. Genel olarak, bütünüyle katı-hal kontakt NHU+ seçici membran elektrotların kondüktometrik sistemde NH4+ iyonuna karşı seçiciliği potansiyometrik sistemle kıyaslandığında azaldı. Ayrıca elektrotların düşük hücre hacimli akış hücreleri hazırlanarak hareketli ortamlarda da ölçümler alındı. D membran bileşimi ile hazırlanan elektrot kullanılarak, amonyumun toprak numunelerinde tayini kondüktometrik ve potansyometrik-akış enjeksiyon sistemlerinde başarıyla gerçekleştirildi.

Özet (Çeviri)

Ill ALL SOLID-STATE CONTACT AMMONIUM-SELECTIVE PVC- MEMBRANE CONDUCTOMETRIC ELECTRODES AND THEIR PERFORMANCE CHARACTERISTICS ABSTRACT All solid-state contact PVC-membrane NH4+-selective electrodes were prepared by using crown-ether compounds Nonactin, dibenzo-18-crown-6 and dicyclo-18-crown-6 as neutral ionofores. The conductometric and potentiometric performances of electrodes were examined with a computerized measuring system. All electrodes almost exhibited Nernstian behaviour (>60 mV/per decade), short response times (<15 s) and long life times (at least 1-2 months) with a linear response between 10“2-10”5 mol.L"1 solutions of primary ion. Selectivity cofficients of the electrodes againt potential interferent ions were calculated by separate solution method. Membrane electrodes A and D exhibited a high degree of conductometric and potentiometric selectivity toward NH/ against Na+ and K+ ions. Although membrane electrodes B and C resulted a limeted selectivity toward NH4+ ion in solution. All solid-state NH/ selective membrane electrodes applied in the potentiometric form of the measurement behaved more selective toward NH4+ than applied in the conductometric form of the measurment selectivity of all solid-state NHU+ selective membrane electrodes toward NH/ ion was diminished by applying conductometry. Flow-through cells for the electrodes were designed and used in flowing conditions. Electrodes prepared with D membrane, ammonium ion was determined successfully in soil samples by the use of conductometric and potentiometric-flow injection system.

Benzer Tezler

  1. Salen bileşenlerinde temel teşkil eden bütünüyle katı hal kontakt pvc membran iyon seçici elektrotlar

    Salen compounds based on all solid-state pvc-membrane ion selective electrodes

    SEMANUR MUTLU

    Yüksek Lisans

    Türkçe

    Türkçe

    2005

    KimyaOndokuz Mayıs Üniversitesi

    Kimya Ana Bilim Dalı

    PROF.DR. İBRAHİM IŞILDAK

  2. Ağır metal seçici elektrotların geliştirilmesi ve hareketli ortamda sülfirid tayini

    Development of heavy metal selective electrodes and determination of sulpiride on the flowing conditions

    ÇİĞDEM YADİGAROĞLU

    Yüksek Lisans

    Türkçe

    Türkçe

    2003

    KimyaOndokuz Mayıs Üniversitesi

    Kimya Ana Bilim Dalı

    DOÇ. DR. İBRAHİM IŞILDAK

  3. Fe+3 ve sülfirid seçici potansiyometrik sensörler

    Fe+3 and sulpiride-sellective potentiometric sensors

    TÜLİN AKGÜN

    Yüksek Lisans

    Türkçe

    Türkçe

    2003

    KimyaOndokuz Mayıs Üniversitesi

    Kimya Ana Bilim Dalı

    DOÇ. DR. İBRAHİM IŞILDAK

  4. Kapiler elektroforez ve potansiyometrik belirleme

    Capillary electrophoresis and potentiometric detection

    EBRU GÜRSOY

    Yüksek Lisans

    Türkçe

    Türkçe

    2007

    KimyaOndokuz Mayıs Üniversitesi

    Kimya Ana Bilim Dalı

    DOÇ.DR. ADEM ASAN

  5. Bütünüyle katı-hal kontak katyon seçici elektrotlar ve Na+-seçici elektrot

    All solid-state contact cation selective electrodes and Na+-selective electrode

    SELİN GÜNDÜZ

    Yüksek Lisans

    Türkçe

    Türkçe

    2006

    KimyaOndokuz Mayıs Üniversitesi

    Kimya Ana Bilim Dalı

    DOÇ. DR. MÜBERRA ANDAÇ