Antirotasyonel mekanizması olan ve olmayan vidalı abutmentların vida gevşemesi ve abutment aşınması açısından in-vitro olarak karşılaştırılması
An in vitro comparison of screw loosening and abutment wearin screw-retained abutments with and without antirotational mechanism
- Tez No: 1013863
- Danışmanlar: PROF. DR. ONUR GEÇKİLİ
- Tez Türü: Doktora
- Konular: Diş Hekimliği, Dentistry
- Anahtar Kelimeler: Protez gevşemesi, Protezler ve implantlar, Prosthesis failure, Prostheses and implants
- Yıl: 2026
- Dil: Türkçe
- Üniversite: İstanbul Üniversitesi
- Enstitü: Sağlık Bilimleri Enstitüsü
- Ana Bilim Dalı: Protetik Diş Tedavisi Ana Bilim Dalı
- Bilim Dalı: Protetik Diş Tedavisi Bilim Dalı
- Sayfa Sayısı: Belirtilmemiş.
Özet
İmplant üstü vidalı sabit protezlerde en sık karşılaşılan mekanik ve teknik komplikasyonlardan biri vida gevşemesidir. Bu in vitro çalışmanın amacı, implant destekli vidalı sabit protetik restorasyonlarda kullanılan antirotasyonel mekanizması olan ve antirotasyonel mekanizması olmayan multi-unit abutmentlarda, okluzal yükleme sonrası oluşan vida gevşemesi ve abutment aşınmasının değerlendirilmesidir. Bu amaçla, antirotasyonel mekanizması olan ve antirotasyonel mekanizması olmayan multi-unit abutmentlar üzerinde kantileverli ve kantileversiz olmak üzere dört farklı grup oluşturulmuş, toplam 40 adet üç üyeli köprü protezi üretilmiştir. Örneklere 1,2 Hz frekansta, 70 N kuvvet altında 240.000 siklus dinamik yükleme uygulanmıştır. Yükleme sonrasında vida gevşeme değerleri dijital tork ölçer ile ölçülmüş, abutment yüzey aşınmaları ise özel yazılım programı kullanılarak değerlendirilmiştir. Kantilever bulunmayan üç üyeli implant destekli köprülerde, vida gevşemesi açısından antirotasyonel mekanizması olan ve antirotasyonel mekanizması olmayan multi-unit abutmentlar arasında anlamlı fark saptanmamıştır (p>0,05). Kantileverli gruplarda ise antirotasyonel mekanizması olan multi-unit abutmentların vida stabilitesi açısından daha avantajlı olduğu belirlenmiştir. Molar bölgede antirotasyonel mekanizmanın abutment aşınma değerlerini artırdığı, premolar bölgede ise gruplar arasında anlamlı fark bulunmadığı saptanmıştır. Tork kaybı ile aşınma arasındaki ilişkinin bölgeye özgü olduğu, özellikle molar bölgede tork kaybı ile aşınma arasında anlamlı ilişki bulunduğu belirlenmiştir (p<0,05). Bu bulgular, posterior bölgede oluşan biyomekanik yüklerin mekanik stabilite ve yüzey kaybı üzerinde birlikte etkili olabileceğini göstermektedir.
Özet (Çeviri)
Screw loosening is reported as one of the frequently encountered mechanical and technical complications in screw-retained fixed implant prostheses. The aim of this in vitro study was to evaluate screw loosening and abutment wear occurring under occlusal loading in multi-unit abutments with and without antirotational mechanisms used in implant-supported screw-retained fixed prosthetic restorations. For this purpose, four different groups were created, and a total of 40 three-unit implant-supported screw-retained fixed prosthetic restorations, with and without cantilevers, were fabricated. A dynamic loading protocol was applied to the specimens at a frequency of 1.2 Hz and a force of 70 N for 240,000 cycles. Screw loosening values were measured, and surface wear analysis was subsequently performed using a spesific software. In three-unit implant-supported screw-retained fixed prosthetic restorations, no significant difference was observed between the groups with and without antirotational mechanisms in terms of screw loosening (p>0.05). In cantilevered three-unit implant-supported screw-retained fixed prosthetic restorations, multi-unit abutments with antirotational mechanisms were more favorable in terms of screw stability. The antirotational mechanism increased abutment wear values in the molar region; however, no significant difference was found between the groups in the premolar region. The relationship between torque loss and wear was found to be site-specific. Specifically, a significant relationship was observed between torque loss and wear in the molar region (p<0.05). These findings suggest that biomechanical loads occurring in the posterior region may affect both mechanical stability and surface loss.
Benzer Tezler
- İntertrokanterik femur kırıklarının tedavisinde proksimal femoral çivi-antirotasyon ve çimentolu kalkar destekli bipolar hemiartroplasti uygulamalarımızın karşılaştırılması
The comparison of outcomes of proximal femoral nail-antirotation and cemented, bipolar hemiarthroplasty with calcar replacement design, in treatment of femur intertrochanteric fractures
HAKAN DUR
Tıpta Uzmanlık
Türkçe
2012
Ortopedi ve TravmatolojiGazi ÜniversitesiOrtopedi ve Travmatoloji Ana Bilim Dalı
PROF. DR. ERDAL CİLA
- Trokanterik bölge femur kırıklarının proksimal femoral çivi antirotasyon (pfna)ile osteosentezi
Başlık çevirisi yok
AHMET İMERCİ
Tıpta Uzmanlık
Türkçe
2010
Ortopedi ve TravmatolojiSağlık BakanlığıOrtopedi ve Travmatoloji Ana Bilim Dalı
DOÇ. DR. LEVENT KARAPINAR
- Proksimal femur çivisi ile tedavi edilen intertrokanterik femur kırığı olgularında cerrahi sonrası abduktor kas güçlerinin shear-wave elastografi ile değerlendirilmesi
The post-op evaluation of the abductor muscle strength by using shear-wave elastography after proximal femur nails treatment in cases with intertrochanteric femur fracture
SÜLEYMAN ALP ÇÖLBE
Tıpta Uzmanlık
Türkçe
2020
Ortopedi ve TravmatolojiTrakya ÜniversitesiOrtopedi ve Travmatoloji Ana Bilim Dalı
DOÇ. DR. MERT ÇİFTDEMİR
- Femur intertrokanterik kırık nedeniyle proksimal femur çivileme yapılan hastaların cerrahi sonrası kalça abdüksiyon gücünün değerlendirilmesi
DUE to intertrochanteric fracture of the femurpatients with proximal femur nailing evaluation of hip abusion strength after surgery
HAKAN ARLI
Tıpta Uzmanlık
Türkçe
2021
Ortopedi ve TravmatolojiSağlık Bilimleri ÜniversitesiOrtopedi ve Travmatoloji Ana Bilim Dalı
DOÇ. DR. ALİ REİSOĞLU
UZMAN İSMAİL ERALP KAÇMAZ
- Köpeklerin değişik derecelerdeki patella luksasyonlarında farklı cerrahi tekniklerin klinik ve radyolojik değerlendirmeler ile karşılaştırılması
Comparision of different surgical techniques in the treatment of canine patellar luxation in various degrees by clinical and radiological evaluation
CENK YARDIMCI
Doktora
Türkçe
2006
Veteriner HekimliğiAnkara ÜniversitesiVeterinerlik Cerrahisi Ana Bilim Dalı
PROF.DR. ARKUN CANDAŞ